N O T I C E
This is only applicable to instruments with two syringes installed.
It is recommended to install the applications (refer to
on page 18) before installing the
titrant. Applications using titrant from syringe 1 are loaded on line 1 of the home screen (refer to
on page 17) and applications using titrant from syringe 2 are loaded on line 2. After the applications are
installed, the correct titrant can be connected to the applicable syringe.
Optional: Fill a desiccant cartridge with an applicable desiccant. Put the desiccant cartridge into the
adapter on the titrant bottle cap. Refer to the illustrated steps that follow, step 1.
Loosen the tube connector on the bottle cap. Push the inlet tube through the connector. Make sure
that the end of the tube is at the bottom of the bottle. Tighten the connector on the bottle cap.
Use the same procedure to connect the second titrant bottle if a second syringe is installed on the
instrument.
Refer to the "Application Note" on the USB applications key to identify the correct pump to connect to
the reagent bottle. Refer to the illustrated steps that follow.