Operation—Foundation Fieldbus
55
6.2.2.6 Cyclic measured value status
The cyclic measured value status is shown in
6.2.2.7 Measured value limits—limit bits
The respective status bit is set when a condition occurs (
status bit is reset when the condition no longer exists.
Table 10 Cyclic measured value status
Priority Quality
Sub-status
Bin-coding
(no limit bits)
Hex-
coding
Low
Good
Good Non-Specific
10 00 00 00
0 x 80
Good Active Advisory Alarm
10 00 10 xx
0 x 88
Good Active Critical Alarm
10 00 11 xx
0 x 8C
Uncertain
Uncertain Non-Specific
01 00 00 xx
0 x 40
Last Usable Value (LUV)
01 00 01 xx
0 x 44
Substitute-Set
01 00 10 xx
0 x 48
Initial Value
01 00 11 xx
0 x 4C
Sensor Conversion Not Accurate 01 01 00 xx
0 x 50
Engineering Unit Violation
01 01 01 xx
0 x 54
Sub-Normal
01 01 10 xx
0 x 58
Bad
Non-Specific
00 00 00 xx
0 x 00
Sensor Failure
00 01 00 xx
0 x 10
Device Value
00 00 11 xx
0 x 0C
High
Out of Service
00 01 11 xx
0 x 1C
Table 11 Limit bit description
Bin coding of limit bits
Description
00
OK
01
Low-limited
10
High-limited
11
Constant
Summary of Contents for si792 T
Page 2: ......
Page 12: ...Specifications 10...
Page 32: ...Installation 30...
Page 36: ...Interface and navigation 34...
Page 66: ...Operation Foundation Fieldbus 64...
Page 80: ...Maintenance 78...
Page 86: ...Troubleshooting 84...
Page 88: ...Replacement parts 86...
Page 94: ...Warranty liability and complaints 92...
Page 98: ...Calibration solutions 96...
Page 106: ...Concentration curves 104...
Page 112: ...ATEX safety instructions 110 Figure 20 Sanitary style sensor...
Page 116: ...Index 114...