SECTION 6,
continued
6.3
Error Codes And Errors
Errors and error codes indicate a functional problem with the
meter and/or the electrode. Error codes numbers will appear in
the temperature field along with other icons. Table 7 describes
the possible error codes, errors, and some possible solutions to
eliminate the cause.
6.4
Meter Service Request Questionnaire
1. What is the complete lot code of the meter and electrode?
2. On what date was the meter purchased?
3. How long has the meter been in use?
4. What types of samples are being tested?
5. What is the temperature of the samples being tested?
6. How often is the meter being used?
7. How is the meter being stored between uses?
8. If the meter has been in use for a while, what maintenance
has been performed?
9. Describe the suspected problem or failure of the meter.
10. Please have your meter, electrode, buffers/standards, and this
completed questionnaire near the phone before calling
technical support.
Table 7 Error Codes
Error Code
Error Name & Display Icons
Description
E2
Cal slope error.
ERROR
and
SLOPE
will be displayed.
Calibration slope is too high or low.
Ensure correct pH buffers are used. Be sure a pH
probe is connected to the meter. Disconnect any
probe connected to the BNC connector.
E3
Cal Std Error.
ERROR, CAL
and
Standard
will be displayed.
The value read will not work in the calibration
algorithm. Repeat calibration with new buffers
or standards.
E9
Corrupt data
Recalled data had a bad checksum. Call Hach
Service.