The default stability drift rate is 0.500 mV/min. Lower stability
criteria (i.e., 0.300 mV/min) will require longer stabilization
times, but the measurement will be more precise. Higher stability
criteria will require shorter stabilization times, but measurements
may be less precise.
To change this option:
1. From the reading mode, press
SETUP
.
2. The
Setup
icon and the number
1
(flashing)
will be displayed.
Press the up arrow until the Setup number is
6
.
3. Change the stability criteria by pressing
ENTER
. The display
will show
_ _ _ _
. The numbered keypad will become active.
4. Enter the desired mV value using the keypad. Press
ENTER
when entry is complete.
5. If a number entry error occurs, start over by pressing
SETUP/CE.
6. When the desired value has been accepted, press
EXIT
to
return to the Reading mode.
3.1.7
Setting the Time
1. From the reading mode, press
SETUP
.
2. The
Setup
icon and the number
1 (flashing)
will be displayed.
Press the up arrow until the Setup number is
7
.
3. Change the time by pressing
ENTER
. The main display will
change to
_ _:_ _,
with the left place holder flashing. The
numerical keypad will become active.
4. Press the desired number key for the left most digit. The
meter uses a 24-hour military clock (12 a.m. = 00:00; 1 p.m.
= 13:00). All four digit places must have a number. If the
left-most digit(s) is not necessary, use zero for the value
(i.e., 08:15 for 8:15 a.m. or 00:30 for 12:30 a.m.).