17
Installation
3.4.3 Wiring for AC Power at the Controller
DANGER
Failure to connect to a good low impedance Protective Earth ground can result in
both a shock hazard and poor performance against electro-magnetic interferences.
1.
Obtain appropriate fittings with IP65 environmental rating.
2.
Remove the display module from the probe module (
).
3.
Remove the four screws securing the probe module front cover. Open the probe
module and disconnect the chassis ground connection from the ground stud to the
cover (
4.
Remove the six screws from the high voltage barrier and remove the barrier.
5.
Insert the wires through the PG1 opening and strain relief fitting or conduit hub.
Tighten the strain relief if used, to secure the cord.
6.
Strip the cable outer insulation 260 mm (10 in.). Shorten all wires except the earth
wire 20 mm (0.78 in.), so the earth cable is 20 mm (0.78 in.) longer than the other
cables (
).
7.
Feed the stripped power cable through the ferrite core twice (
) and wire into
. Tug gently after each insertion to
ensure the connection is secure.
8.
Seal any unused openings in the controller box with conduit opening sealing plugs.
9.
Reinstall the high voltage barrier.
10. Reconnect the chassis ground connection to the ground stud of the probe
module cover.
11. Reinstall the probe module cover and screw into place.
Figure 12
Proper Wire Preparation and Ferrite Core Wiring
1.
Power Cable Wire Preparation
3.
Power Cable Wires
2.
Ferrite Core
4.
Power Cables Wrapped Around Ferrite Core
2
4
3
1
260 mm (10 inches)
240 mm (9.75 inches)
客服:400 688 9809
网址:www.mamots.com 电话:0431-85178766
Summary of Contents for sc1000
Page 1: ...Visit http www hach com 400 688 9809 www mamots com 0431 85178766...
Page 38: ...Visit us at www hach com 400 688 9809 www mamots com 0431 85178766...
Page 58: ...Visit us at www hach com 400 688 9809 www mamots com 0431 85178766...
Page 62: ...Visit us at www hach com 400 688 9809 www mamots com 0431 85178766...