19
Installation
Figure 15
Inside the AC Probe Module
3.4.4 Installing Power Outlets
Important Note: The AC power outlets cannot be used with the 24 VDC power supply.
AC power outlets can be installed to connect AC powered sc-sensors to the sc1000.
Refer to
and the following directions for proper power socket installation.
1.
Remove power from the instrument.
2.
Remove the probe module cover and remove the voltage barrier cover.
3.
Remove the plug from the power socket housing opening. Feed the power socket
cable through the housing opening.
Table 5 AC Power Wiring Information
Terminal Number
Terminal Description
Wire Color Code for North America
L
Hot (L1)
Black
N
Neutral (N)
White
Protective Earth (PE)
Green
1.
Fan
8.
AC power connections
2.
Main circuit board
9.
Earth ground connection
3.
Connector for expansion slot #2
10. Power outlet connection
4.
Connector for expansion slot #3
11. Connector for expansion slot #1
5.
Connector for expansion slot #4
12. Sensor connections
6.
Fuse (2), F1 and F2: M 3.5A, medium slow-low
13. Relay card connection
7.
Fuse (2), F3 and F4: T 8A; 100–240 V
F2
F1
F4
F3
1
L
N
2
3
4
3
2
1
4
5
6
8
7
12
13
11
9
10
客服:400 688 9809
网址:www.mamots.com 电话:0431-85178766
Summary of Contents for sc1000
Page 1: ...Visit http www hach com 400 688 9809 www mamots com 0431 85178766...
Page 38: ...Visit us at www hach com 400 688 9809 www mamots com 0431 85178766...
Page 58: ...Visit us at www hach com 400 688 9809 www mamots com 0431 85178766...
Page 62: ...Visit us at www hach com 400 688 9809 www mamots com 0431 85178766...