Operation
5.4.2 Current Inputs Setup
Select INPUT INT
Select INPUT 1–4
EDIT NAME
Allows user to enter the name of the input device.
DEVICE NAME
Name of the device on the current input.
PARAMETER NAME
Abbreviation for the parameter measured.
SET PARAMETER
Select the appropriate displayed parameter.
SELECT UNITS
Select the units to be displayed.
SET FUNCTION
Select ANALOG or DIGITAL for the type of sensor being used.
SET FILTER
Select the reaction time.
SCALE 0mA/ 4mA
Select 0 mA or 4 mA for minimum current (outputs will be set to 0–20 mA or 4–20 mA).
SET LOW VALUE
This value is to be displayed by the sc1000 controller system at the lowest measured value (in the unit for the measured
value for the probe). This is the 0mA or 4 mA value.
SET HIGH VALUE
This value is to be displayed by the sc1000 controller system at the highest measured value (in the unit for the measured
value for the probe). This is the 20 mA value.
ON ERROR MODE
Select between HOLD and SET TRANSFER. During calibration, the input can remain active, held, or transferred to a
preset mA value.
CONCENTRATION
Current mA input value from the sensor.
VERSION—The version of the software is displayed.
LOCATION—Information on the location of the device can be edited.
客服:400 688 9809
网址:www.mamots.com 电话:0431-85178766
Summary of Contents for sc1000
Page 1: ...Visit http www hach com 400 688 9809 www mamots com 0431 85178766...
Page 38: ...Visit us at www hach com 400 688 9809 www mamots com 0431 85178766...
Page 58: ...Visit us at www hach com 400 688 9809 www mamots com 0431 85178766...
Page 62: ...Visit us at www hach com 400 688 9809 www mamots com 0431 85178766...