![Hach POLYMETRON 8810 ISE Basic User Manual Download Page 3](http://html.mh-extra.com/html/hach/polymetron-8810-ise/polymetron-8810-ise_basic-user-manual_577099003.webp)
Specifications
Specifications are subject to change without notice.
Specification
Details
Ambient temperature
5 - 40 °C (41 - 104 °F)
Relative humidity
10 to 80%
Operating altitude
From 0 to 2,000 m. (6,550 ft.) above sea level
Mains power supply
110/220/240 VAC, 50/60 Hz, ± 10%
Fuses: 110 to 120V, T630mAL250V; 220V to 240V, T1.25AL250V
Max. consumption
110 VA
Overvoltage category
2 (according to standard EN 61010-1)
Pollution degree
2
CE compliance
EN61326-1: EMC Directive
Note: This is a Class A product. In a domestic environment this product may cause radio
interference in which case the user may be required to take adequate measures.
EN61010-1: LVD Directive
ETL approved
ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
Korean certification
User Guidance for EMC Class A Equipment
업무용을
위한
EMC
등급
A
장치에
대한
사용자
지침
사용자안내문
A
급
기기
(
업무용
방송통신기자재
)
이
기기는
업무용
(A
급
)
전자파적합기기로서
판매자
또는
사용자는
이
점을
주의
하시기
바라며
,
가정외의
지역에서
사용하는
것을
목적으로
합니다
.
Compressed air
4 - 7 bar, filtered and dried
Rinse water pressure
0.5 - 6 bar
Reagents
10 liter storage canister (supplied)
Analog outputs
Number: 2; 0 - 20 or 4 - 20 linear programmable
Alarms
2 x Relay (concentration); 1 x Warning or System alarm
Exchangeable sensors
pH + reference electrode
Temperature compensation
Pt100
Level control
Sample; Reagents; Calibration solution; Chemical cleaning
Pumps
Peristaltic, micro piston, pulse or volumetric pumps for calibration and conditioning
reagents
Tubing
Tygon
®
; Polyethylene
Reagent consumption
Application specific
Titration time
Application specific
Cycle time
Programmable up to 999 minutes
Accuracy
< ± 2 to 4% (application specific)
Reproducibility
< ± 2 to 4% (application specific)
English
3
Summary of Contents for POLYMETRON 8810 ISE
Page 2: ...English 3 Deutsch 27 Italiano 53 Français 79 Español 104 Polski 130 Română 156 2 ...
Page 180: ......