Table 3  Cleaning agents

Contaminant

Cleaning agent

Time

Grease, oils and fats

Electrode cleaning solution

2 hours maximum

Mineral buildup

10% hydrochloric acid (HCl) solution

5 minutes maximum

Replace the batteries

When the battery icon flashes or the tester will not come on, replace all four batteries.
Refer to 

Install the batteries

 on page 5.

Troubleshooting

Message

Possible cause

Solution

"- - - -"
(top line)

The parameter value is not within the
measurement range of the tester. Refer to 

Specifications

 on page 2.

Make sure that the value of the sample is within the
measurement range of the tester. Gently clean the
sensor. Refer to 

Clean the sensor

 on page 8. Calibrate

as necessary.

"- - - -"
(bottom line)

The temperature value is not within the
operating temperature range of the tester or
a temperature sensor failure occurred. Refer
to 

Specifications

 on page 2 for the operating

temperature range.

Make sure that the sample temperature is within the
operating temperature range of the tester. Contact
technical support as necessary.

Note: The tester can still be used if a temperature sensor failure
has occurred, but without automatic temperature compensation.

Battery icon flashes The batteries have less than 10% power

remaining.

Replace all four batteries. Refer to 

Install the batteries

on page 5.

Replacement parts and accessories

W A R N I N G

 

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to
the instrument or equipment malfunction. The replacement parts in this section are
approved by the manufacturer.

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate

distributor or refer to the company website for contact information.

Table 4  Replacement parts

Description

Quantity

Item no.

AAA alkaline batteries

4/pkg

4674300 

Lanyard

201305 

SINGLET, for LR tester, 1413 µS/cm

20/pkg

2771420 

SINGLET, for HR tester, 12.88 mS/cm

20/pkg

2771520 

Table 5  Accessories

Description

Quantity

Item no.

SINGLET, for LR tester, 147 µS/cm

20/pkg

2771320 

Electrode cleaning solution

500 mL

2965249 

Hydrochloric Acid (HCl)

2.5 L

13406 

Hydrochloric Acid (HCl)

500 mL

13449 

9

Summary of Contents for Pocket Pro CondLR

Page 1: ...User Manual DOC022 52 80396 Pocket Pro CondLR CondHR 1...

Page 2: ...equential damages resulting from any defect or omission in this manual The manufacturer reserves the right to make changes in this manual and the products it describes at any time without notice or ob...

Page 3: ...plier for instructions on how to return end of life equipment producer supplied electrical accessories and all auxiliary items for proper disposal Certification Canadian Radio Interference Causing Equ...

Page 4: ...tion the receiving antenna for the device receiving the interference 3 Try combinations of the above Product overview The Pocket Pro conductivityLR tester and conductivityHR tester measure the conduct...

Page 5: ...ical type and are inserted in the correct orientation Do not mix new and used batteries WA RNING Fire hazard Battery substitution is not permitted Use only alkaline batteries Install the four AAA alka...

Page 6: ...y power is less than 10 Shows when the lock feature is on When the lock feature is on the parameter value on the display does not change Push to set the lock feature to on or off Shows when the tester...

Page 7: ...standard CondLR tester One or two auto recognition calibration standards 1 Set the power to on 2 Remove the cap from the sensor 3 Push to go to calibration mode The auto recognition standard s the te...

Page 8: ...perature compensation factor linear 0 00 to 9 99 per C default 2 00 per C The correction factor may need to be identified experimentally For example the factor for ultrapure water is 4 55 per C and th...

Page 9: ...rature range of the tester Contact technical support as necessary Note The tester can still be used if a temperature sensor failure has occurred but without automatic temperature compensation Battery...

Page 10: ...ach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Com...