Table 2  Key functions

Key

Description

Push and hold to set the power to on or off.

Push to start a calibration. To exit a calibration, push and hold.
Push and hold until "SEt" shows on the display to go to the settings menu. To exit the settings menu, push and hold
until "End" shows on the display. When in the settings menu, push to scroll through the settings.

Note: Power cannot be set to off while in settings or calibration mode.

Push to set the lock feature to on or off. When the lock feature is on, the lock icon shows and the parameter value on
the display does not change.

Calibration

C A U T I O N

 

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.

Calibrate the tester before initial use and when:

Results drift

Results are not accurate

Calibrate

Items to collect:

 Cond

HR

 tester: One auto-recognition calibration standard, Cond

LR

tester: One or two auto-recognition calibration standards

1.

Set the power to on.

2.

Remove the cap from the sensor.

3.

Push   to go to calibration mode.
The auto-recognition standard(s) the tester expects to measure shows on the bottom
line.

4.

Rinse the sensor and cap with deionized water and blot dry.

5.

Pour the calibration standard shown into the cap to the fill line.

6.

Put the sensor fully into the cap.

7.

When the measurement is stable, push   to save the calibration and go to
continuous measurement mode. The measured value will flash 3 times and then stop.
Then, "END" shows on the display.

8.

(Optional) For the LR tester, to measure a second calibration standard, do steps 

4

7

again.

9.

Rinse the sensor and cap with deionized water and blot dry.

Measurement

Note: Air bubbles under the probe tip when submerged can cause slow stabilization or error in

measurement. Shake the tester from side to side to remove air bubbles.

1.

Set the power to on.

2.

Remove the cap from the sensor.

3.

If the lock icon shows on the display, push   to go to continuous measurement mode.

4.

Rinse the sensor and cap with deionized water and blot dry.

5.

Pour the water sample into the cap to the fill line.

7

Summary of Contents for Pocket Pro CondLR

Page 1: ...User Manual DOC022 52 80396 Pocket Pro CondLR CondHR 1...

Page 2: ...equential damages resulting from any defect or omission in this manual The manufacturer reserves the right to make changes in this manual and the products it describes at any time without notice or ob...

Page 3: ...plier for instructions on how to return end of life equipment producer supplied electrical accessories and all auxiliary items for proper disposal Certification Canadian Radio Interference Causing Equ...

Page 4: ...tion the receiving antenna for the device receiving the interference 3 Try combinations of the above Product overview The Pocket Pro conductivityLR tester and conductivityHR tester measure the conduct...

Page 5: ...ical type and are inserted in the correct orientation Do not mix new and used batteries WA RNING Fire hazard Battery substitution is not permitted Use only alkaline batteries Install the four AAA alka...

Page 6: ...y power is less than 10 Shows when the lock feature is on When the lock feature is on the parameter value on the display does not change Push to set the lock feature to on or off Shows when the tester...

Page 7: ...standard CondLR tester One or two auto recognition calibration standards 1 Set the power to on 2 Remove the cap from the sensor 3 Push to go to calibration mode The auto recognition standard s the te...

Page 8: ...perature compensation factor linear 0 00 to 9 99 per C default 2 00 per C The correction factor may need to be identified experimentally For example the factor for ultrapure water is 4 55 per C and th...

Page 9: ...rature range of the tester Contact technical support as necessary Note The tester can still be used if a temperature sensor failure has occurred but without automatic temperature compensation Battery...

Page 10: ...ach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Com...

Reviews: