11
Introduction,
continued
electrode body and rotate the cartridge clockwise until it is
seated (Figure 2).
2. Place the dispenser unit over the electrolyte cartridge.
Screw the dispenser unit onto the electrode body until it
stops turning. Do not over tighten.
Figure 2
Assembling the Platinum Series Electrode
Note:
The electrode will be automatically primed by screwing the dispenser unit
onto the electrode body.
3. If electrolyte gel is not visible at the reference outlet,
depress the pump button until it clicks; release the button.
Repeat this procedure until gel is visible at the reference
outlet. Alternatively, fully depress the button and rotate it
clockwise until gel is visible at the reference outlet (1 to 3
rotations).