74
Plumbing and connection options
A.9 Option 5 plumbing and connections
Option 5 uses an sc analyzer as a 2-channel analyzer with two
FILTRAX (FILTRAX 1 and FILTRAX 2), supplying two continuous
sample streams. The waste from the analyzer and both FILTRAX is
discharged back into the drain through the optional heated drain
hose LZY302 (230 V) or LZY303 (115 V).
Refer to
and the following instructions for Option 5:
1.
Install both FILTRAX into the sample stream. Refer to the
FILTRAX User Manual for more information.
2.
Feed the heated hose from FILTRAX 1 through the analyzer
opening (
, item 7). Use Seal Plug #1 to secure.
3.
Feed the heated hose from FILTRAX 2 through the analyzer
opening (item 6). Use Seal Plug #1 to secure.
4.
Feed the heated drain hose through the analyzer (item 5). Use
Seal Plug #1 to secure.
Note:
The two samples lines of the heated drain hose are not used.
5.
Connect the heated drain power connections. Refer to
6.
Connect the heated drain tube to the T-fitting (item 8).
7.
Connect the sample line from FILTRAX 1 to the bottom inlet on
Overflow Vessel 1 using the fittings (item 11).
8.
Connect the sample line from FILTRAX 2 to the bottom inlet on
Overflow Vessel 2 using the fittings (item 12).
Summary of Contents for PHOSPHAX indoor sc
Page 2: ......
Page 6: ...6 Specifications Figure 1 Instrument dimensions PHOSPHAX sc...
Page 8: ...8 Specifications...
Page 34: ...34 System Startup...
Page 55: ...55 Replacement Parts and Accessories Figure 21 Analyzer enclosure PHOSPHAX indoor sc...
Page 56: ...56 Replacement Parts and Accessories Figure 22 Analyzer panel overview...
Page 57: ...57 Replacement Parts and Accessories Figure 23 Analyzer panel front view details...
Page 58: ...58 Replacement Parts and Accessories Figure 24 Analyzer panel back view details...
Page 62: ...62 Limited Warranty...