Hach MTC10101 User Instructions Download Page 9

DOC272.53.80033

9

Maintenance

Clean the probe if the response time becomes slow (

Clean the probe on page 9

).

Storage

Note: For optimal performance, the reference junction must not dry out.

Short-term storage:

Store the electrode in 

Hach Electrode Storage Solution

 or 

3M potassium chloride 

(KCl) solution

. The electrode can also be stored for up to 2 hours in a sample (if the 

sample pH is not extreme).

Long-term storage:

1.

Rinse the electrode with deionized water. Dry the electrode with a lint-free cloth.

2.

Fill the electrode storage cap halfway with 

Hach Electrode Storage Solution

 or 

3M 

potassium chloride (KCl) solution

.

3.

Install the electrode storage cap. Make sure that the storage cap covers the reference 
junction completely.

Rugged probe maintenance

The shroud protects the sensor elements during rugged applications. Damage to the 
sensing elements can occur if the shroud is not installed during field use. Damage under 
these conditions is not covered under warranty.

To remove the shroud:

1.

Unscrew the locking ring and slide the shroud and locking ring off the probe.

2.

Slide the shroud and locking ring off the probe.

To install the shroud

1.

Place the locking ring on the probe body with the threads toward the platinum disc.

2.

Slide the shroud on the probe until it is seated in the locking groove.

3.

Hand-tighten the locking-ring on the shroud.

Clean the probe

For general cleaning (including oils, greases, and organics):

Rinse the electrode with deionized water and blot dry with a lint-free cloth.

Immerse the sensor array and platinum disc in Electrode Cleaning Solution or 
warm detergent solution for up to 15 minutes.

Note: The platinum disc can be polished using a soft cloth or cotton swab with detergent 
solution.

Rinse the electrode sensor array and platinum disc with deionized water. Blot dry 
with a lint-free cloth.

For inorganic deposits

Immerse the platinum disc in a solution of 0.1M hydrochloric or nitric acid solution 
and soak for up to 15 minutes.

Rinse the electrode sensor array and platinum disc with deionized water. Blot dry 
with a lint-free cloth.

After cleaning has been performed, condition the platinum electrode in 
representative sample prior to use.

 

Summary of Contents for MTC10101

Page 1: ...EC European electrical equipment users must now return old or end of life equipment to the Producer for disposal at no charge to the user Note For return for recycling please contact the equipment pr...

Page 2: ...int free cloth 4 If the electrode will not be used immediately fill the storage cap halfway with Hach Electrode Storage Solution or 3M KCl and install the cap 5 Install the shroud and locking ring and...

Page 3: ...be must have the correct service life time stamp Set the date and time in the meter before attaching the probe It is not necessary to recalibrate when moving a probe from one HQd meter to another if t...

Page 4: ...er blot dry and place it in the standard solution Make sure that the sample level is above the reference junction 5 Push Read Stir gently The display will show Stabilizing and a progress bar as the pr...

Page 5: ...estion mark until the electrode is recalibrated Recalibrate the electrode to correct the status indicator Calibration on page 3 Measurement Measurement notes Data is stored automatically in the data l...

Page 6: ...nect the electrode to the meter 3 Rinse the electrode with the sample and place the electrode in fresh sample 4 Push Read The display will show Stabilizing and a progress bar as the probe stabilizes i...

Page 7: ...ng many substances in solution A redox titration consists of adding to an unknown sample small increments of a titrant that converts the unknown to a different oxidation state After each addition of t...

Page 8: ...ng the potential after each addition Near the end point when large potential changes appear add increments of 0 1 0 2 mL Continue the titration 3 4 mL past the end point 5 Plot the electrode potential...

Page 9: ...d 1 Unscrew the locking ring and slide the shroud and locking ring off the probe 2 Slide the shroud and locking ring off the probe To install the shroud 1 Place the locking ring on the probe body with...

Page 10: ...ings which cannot be changed A saved method can be used instead of repeatedly adjusting individual settings Changes made to a user defined method are automatically saved with the existing name Multipl...

Page 11: ...V The measurement limits can be set to match the acceptable values for the sample When the measurement is above the upper limit setting or below the lower limit setting the meter will show an Out of l...

Page 12: ...200 0 mV Custom standard value Custom standards are characterized at 25 C Table 4 Calibration options settings Settings Option Description Table 5 Check standard options settings Settings Options Desc...

Page 13: ...sponse time Contaminated reference junctions Rinse reference junction with deionized water thoroughly shake the electrode downward to remove any air bubbles and soak in 3 0 M KCl storage solution for...

Page 14: ...ts Check standard value is outside of limits set in the current method Verify that the standard is within the limits of the current method Create a new method that expands the acceptable limits Measur...

Page 15: ...storage cap lid 1 5829800 Field kit Includes glove kit and 5 120 mL sample cups 1 5825800 Glove kit only for HQd meters 1 5828700 IntelliCAL probe stand for standard probes 1 8506600 Meter stand 1 850...

Page 16: ...insert for meter and two probe storage cavities space for storage of HQd meter with protective glove 1 8505600 Rugged shroud kit 1 5825900 Consumables Electrode cleaning solution 500 mL 2965249 Electr...

Reviews: