Hach MTC10101 User Instructions Download Page 7

DOC272.53.80033

7

Measurement - conversion to SHE reference

For some applications it is customary to report redox potential readings relative to the 
standard hydrogen electrode (SHE), also called normal hydrogen electrode (NHE). To do 
this, select the value in 

Table 1  Standard potential of reference electrode

 that 

corresponds to the temperature of the solution measured. Substitute that value E

ref

 into 

the equation and solve for E

h

:

E

h

 = E + E

ref

where:

E

= oxidation reduction potential of the sample relative to the SHE

E = potential developed by the ORP/Redox electrode
E

ref 

= potential developed by the reference electrode portion relative to the SHE (refer to 

Table 1  Standard potential of reference electrode

).

Table 1

 shows the potentials, E

ref

, developed by the reference electrode portion relative to 

the SHE at various temperatures.

Measurement - Oxidation Titrations

Oxidation-reduction, or redox titrations, provide a simple, reliable method for determining 
many substances in solution. A redox titration consists of adding to an unknown sample 
small increments of a titrant that converts the unknown to a different oxidation state. After 
each addition of titrant, the ORP/Redox electrode develops a potential proportional to the 
logarithm of the ratio of the activities of the two oxidation states.

At the inflection, or end point, the titrant has completely oxidized or reduced the unknown, 
causing a sharp change in the logarithm of the ratio of the activities of the two oxidation 
states. A corresponding sharp change in the potential is developed by the platinum 
electrode. Often several oxidizing or reducing species can be precisely determined in the 
same solution by a single titration with several inflection points.

The following directions represent a general procedure for performing a redox titration 
once the sample is prepared for measurement.

Table 1 Standard potential of reference electrode

Temperature (ºC)

Electrode Potential in mV (E

ref

)

80

163.1

75

167.7

70

172.1

65

176.4

60

180.3

55

184.4

50

188.4

45

192.3

40

196.1

35

199.8

30

203.4

25

207.0

20

210.5

15

214.0

10

217.4

5

220.9

0

224.2

Summary of Contents for MTC10101

Page 1: ...EC European electrical equipment users must now return old or end of life equipment to the Producer for disposal at no charge to the user Note For return for recycling please contact the equipment pr...

Page 2: ...int free cloth 4 If the electrode will not be used immediately fill the storage cap halfway with Hach Electrode Storage Solution or 3M KCl and install the cap 5 Install the shroud and locking ring and...

Page 3: ...be must have the correct service life time stamp Set the date and time in the meter before attaching the probe It is not necessary to recalibrate when moving a probe from one HQd meter to another if t...

Page 4: ...er blot dry and place it in the standard solution Make sure that the sample level is above the reference junction 5 Push Read Stir gently The display will show Stabilizing and a progress bar as the pr...

Page 5: ...estion mark until the electrode is recalibrated Recalibrate the electrode to correct the status indicator Calibration on page 3 Measurement Measurement notes Data is stored automatically in the data l...

Page 6: ...nect the electrode to the meter 3 Rinse the electrode with the sample and place the electrode in fresh sample 4 Push Read The display will show Stabilizing and a progress bar as the probe stabilizes i...

Page 7: ...ng many substances in solution A redox titration consists of adding to an unknown sample small increments of a titrant that converts the unknown to a different oxidation state After each addition of t...

Page 8: ...ng the potential after each addition Near the end point when large potential changes appear add increments of 0 1 0 2 mL Continue the titration 3 4 mL past the end point 5 Plot the electrode potential...

Page 9: ...d 1 Unscrew the locking ring and slide the shroud and locking ring off the probe 2 Slide the shroud and locking ring off the probe To install the shroud 1 Place the locking ring on the probe body with...

Page 10: ...ings which cannot be changed A saved method can be used instead of repeatedly adjusting individual settings Changes made to a user defined method are automatically saved with the existing name Multipl...

Page 11: ...V The measurement limits can be set to match the acceptable values for the sample When the measurement is above the upper limit setting or below the lower limit setting the meter will show an Out of l...

Page 12: ...200 0 mV Custom standard value Custom standards are characterized at 25 C Table 4 Calibration options settings Settings Option Description Table 5 Check standard options settings Settings Options Desc...

Page 13: ...sponse time Contaminated reference junctions Rinse reference junction with deionized water thoroughly shake the electrode downward to remove any air bubbles and soak in 3 0 M KCl storage solution for...

Page 14: ...ts Check standard value is outside of limits set in the current method Verify that the standard is within the limits of the current method Create a new method that expands the acceptable limits Measur...

Page 15: ...storage cap lid 1 5829800 Field kit Includes glove kit and 5 120 mL sample cups 1 5825800 Glove kit only for HQd meters 1 5828700 IntelliCAL probe stand for standard probes 1 8506600 Meter stand 1 850...

Page 16: ...insert for meter and two probe storage cavities space for storage of HQd meter with protective glove 1 8505600 Rugged shroud kit 1 5825900 Consumables Electrode cleaning solution 500 mL 2965249 Electr...

Reviews: