Drifting/Inaccurate 
readings

Contaminated platinum disc

Clean the electrode.

Clogged reference

Rinse reference junction holes with deionized water 
thoroughly and gently tap the electrode with hand or 
shake the electrode down to remove any air bubbles.

Improper storage conditions 

Clean or condition the electrode and attempt 
recalibration.The electrode may not function correctly if 
the electrode has been left dry for extended periods of 
time. Attempt to recondition the reference junctions by 
soaking the electrode tip in a 3.0 M KCl storage solution 
for 1-2 hours. Rinse with deionized water prior to use.

Electromagnetic Forces (EMF) such as 
voltaic cells, thermoelectric devices, 
electrical generators, resistors and 
transformers.

Avoid testing in areas where EMF is present. For testing 
in process units (i.e. spot checking), make sure the 
equipment is grounded.

Air bubbles around inner reference 
electrode

Gently tap the electrode with hand or shake the electrode 
downward to remove any air bubbles

Out of limits

Check standard value is outside of limits 
set in the current method

Verify that the standard is within the limits of the current 
method.

Create a new method that expands the acceptable limits.

Measurement value is outside of the 
measurement limits set in the current 
method.

Verify that the sample is within the limits of the current 
method.

Create a new method with an expanded range. 

Calibration adjustment offset value 
outside the limits set in the current 
method.

Verify that the standard is within the limits of the current 
method. 

Create another method that expands the acceptable 
limits.

Storage cap not removed

Remove storage cap

Temperature out of 
range

Calibration temperature value is outside 
of range

Verify that the sample temperature is within the range of 
the electrode.

Verify that the temperature sensor is working correctly.

Measured temperature is outside range 
of the electrode

Verify that the standard temperature is within the range of 
the electrode.

Verify that the temperature sensor is working correctly.

Check standard temperature value is 
outside of range

Verify that the Check Standard temperature is within the 
range of the electrode.

Troubleshooting (continued)

Message or 
symptom

Possible cause

Action

Summary of Contents for MTC10101

Page 1: ...EC European electrical equipment users must now return old or end of life equipment to the Producer for disposal at no charge to the user Note For return for recycling please contact the equipment pr...

Page 2: ...int free cloth 4 If the electrode will not be used immediately fill the storage cap halfway with Hach Electrode Storage Solution or 3M KCl and install the cap 5 Install the shroud and locking ring and...

Page 3: ...be must have the correct service life time stamp Set the date and time in the meter before attaching the probe It is not necessary to recalibrate when moving a probe from one HQd meter to another if t...

Page 4: ...er blot dry and place it in the standard solution Make sure that the sample level is above the reference junction 5 Push Read Stir gently The display will show Stabilizing and a progress bar as the pr...

Page 5: ...estion mark until the electrode is recalibrated Recalibrate the electrode to correct the status indicator Calibration on page 3 Measurement Measurement notes Data is stored automatically in the data l...

Page 6: ...nect the electrode to the meter 3 Rinse the electrode with the sample and place the electrode in fresh sample 4 Push Read The display will show Stabilizing and a progress bar as the probe stabilizes i...

Page 7: ...ng many substances in solution A redox titration consists of adding to an unknown sample small increments of a titrant that converts the unknown to a different oxidation state After each addition of t...

Page 8: ...ng the potential after each addition Near the end point when large potential changes appear add increments of 0 1 0 2 mL Continue the titration 3 4 mL past the end point 5 Plot the electrode potential...

Page 9: ...d 1 Unscrew the locking ring and slide the shroud and locking ring off the probe 2 Slide the shroud and locking ring off the probe To install the shroud 1 Place the locking ring on the probe body with...

Page 10: ...ings which cannot be changed A saved method can be used instead of repeatedly adjusting individual settings Changes made to a user defined method are automatically saved with the existing name Multipl...

Page 11: ...V The measurement limits can be set to match the acceptable values for the sample When the measurement is above the upper limit setting or below the lower limit setting the meter will show an Out of l...

Page 12: ...200 0 mV Custom standard value Custom standards are characterized at 25 C Table 4 Calibration options settings Settings Option Description Table 5 Check standard options settings Settings Options Desc...

Page 13: ...sponse time Contaminated reference junctions Rinse reference junction with deionized water thoroughly shake the electrode downward to remove any air bubbles and soak in 3 0 M KCl storage solution for...

Page 14: ...ts Check standard value is outside of limits set in the current method Verify that the standard is within the limits of the current method Create a new method that expands the acceptable limits Measur...

Page 15: ...storage cap lid 1 5829800 Field kit Includes glove kit and 5 120 mL sample cups 1 5825800 Glove kit only for HQd meters 1 5828700 IntelliCAL probe stand for standard probes 1 8506600 Meter stand 1 850...

Page 16: ...insert for meter and two probe storage cavities space for storage of HQd meter with protective glove 1 8505600 Rugged shroud kit 1 5825900 Consumables Electrode cleaning solution 500 mL 2965249 Electr...

Reviews: