Modbus register maps
A.3 Data label
provides a register for sample and analog data labels.
A.4 Sample data
Sample data records (
) are updated at each polled interval regardless of the
sample and hold times in the configuration registers. If real-time data is not required, use
the buffered records (address 500+).
Table 18 Count bin data labels
Address
Register description
Access
Size (bytes) Data format
200
Size 1 label
R/P
4
0.2–10.0 microns
202
Size 2 label
R/P
4
0.2–10.0 microns
204
Size 3 label
R/P
4
0.2–10.0 microns
206
Size 4 label
R/P
4
0.2–10.0 microns
208–231
Reserved
232
Analog input 1 label
R
4
CAL
234
Analog input 2 label
R
4
TMP
236
Analog input 3 label
R
4
RH
238
Analog input 4 label
R
4
FLO
240–251
Reserved
252–299
Expansion
Table 19 Sample data
Address
Register description
Access
Size (bytes) Notes
300
Sample UTC timestamp—year
R
2
YYYY (2000–9999)
301
Sample UTC timestamp—month/day
R
2
MD (1–12, 1–31)
302
Sample UTC timestamp—hour
R
2
H (0–23)
303
Sample UTC timestamp—minute/second
R
2
MS (0–59, 0–59)
304
Sample period—hours
R
2
H (0-23)
305
Sample period—minutes/seconds
R
2
MS (0–59, 0–59)
306
Reserved
307-308
Sample volume
R
4
309
Sample status. Refer to
.
R
2
Bit wise mapped
310
Reserved
311
Size 1 counts
R
4
0–4,294,967,295
313
Size 2 counts
R
4
0–4,294,967,295
315
Size 3 counts
R
4
0–4,294,967,295
317
Size 4 counts
R
4
0–4,294,967,295
319–342
Reserved
343
Analog channel 1 (flow)
R
2
mV
344
Analog channel 2 (temperature)
R
2
0.1 ºC external probe only
345
Analog channel 3 (relative humidity)
R
2
0.1% RH external probe only
346
Analog channel 4
R
2
347
Analog channel 5 (CAL)
R
2
mV
348–352
Reserved
353
Location name
R
32
Double byte characters (16)
Summary of Contents for Met One 6000 Series
Page 2: ......
Page 33: ...31 Operation Figure 21 Setup utility program non wireless units...
Page 48: ...46 Troubleshooting...
Page 50: ...48 Replacement parts and accessories...
Page 52: ...50 Contact information...
Page 54: ...52 Limited warranty...
Page 64: ...62 Modbus register maps...