16
Installation
3.4.3 Sample probe installation
The sample probe must be installed correctly to prevent contamination of the counter and
to get a representative sample of the area.
3.4.3.1 Sample probe kits
The following optional kits are available for installing the sampling probe. Refer to
Parts and accessories on page 47
•
Direct mount—the probe is installed on a short piece of tubing directly on top of the
counter (
). Use this probe when the particle counter can be located where the
sample is collected. Use the direct mount probe to keep particle loss to a minimum.
•
T-type wall bracket—the probe is installed in a wall bracket. The tubing is cut to
connect the probe to the counter.
•
Vertical wall mount—the probe is connected to a stainless steel tube and bracket. Use
this probe for installation on equipment with stainless steel tubing. The probe can be
located where the sample is collected.
•
Through-wall mount—the probe is connected to a stainless steel tube and wall
bracket. Use this probe to collect samples on the other side of a wall from the counter.
•
Ceiling mount—the probe is connected to a stainless steel tube (J-hook or 90 degree)
and ceiling bracket. Use this probe to collect samples in the middle of a room or when
the particle counter is installed above the room.
Figure 9 Probe mount options
1
Through-wall mount
4
Vertical wall mount
2
Ceiling mount—J hook
5
T-type wall bracket
3
Ceiling mount—90 degree
Summary of Contents for Met One 6000 Series
Page 2: ......
Page 33: ...31 Operation Figure 21 Setup utility program non wireless units...
Page 48: ...46 Troubleshooting...
Page 50: ...48 Replacement parts and accessories...
Page 52: ...50 Contact information...
Page 54: ...52 Limited warranty...
Page 64: ...62 Modbus register maps...