42
Maintenance
6.2.2 Clean the instrument enclosure
With the enclosure securely latched, use a soft cloth and a mild detergent to wipe the
outside of the enclosure. Do not allow moisture to enter the enclosure.
6.2.3 Clean the colorimeter
The colorimeter measuring cell may collect sediment or develop a film growth on the
inside walls. Cleaning with an acid solution and a cotton swab on a monthly basis is
recommended. Depending on sample conditions, it may be necessary to clean the cell
more often.
CAUTION
Chemical exposure hazard. The chemicals used in this procedure may be
hazardous if inappropriately handled or accidentally misused. Please read all
warnings on the reagent labels. Protective eye wear is always recommended when
contact with chemicals is possible.
Colorimeter cell cleaning procedure:
1.
Press the
MENU
key, then press the down
ARROW
key until the
MAINT
menu
appears. Press
ENTER
.
2.
Press the down
ARROW
key to scroll to the
CLEAN
option. Press
ENTER
.
3.
Refer to
Figure 22
and locate the colorimeter.
4.
Remove the rubber plug on the top of the colorimeter assembly.
5.
Remove the magnetic stir bar using a bent steel paper clip as a retrieval tool. Refer
to
Figure 22
.
6.
Wait until
CLEAN
flashes on the status line of the display, then fill the colorimeter with
19.2 N Sulfuric Acid Standard Solution (203832).
Note: Lower normality sulfuric acid will NOT be sufficient for cleaning the colorimeter.
7.
Allow the sulfuric acid to stand in the colorimeter for 15 minutes.
8.
Insert a wooden or paper cotton-tipped swab
*
into the opening and move it up and
down with a gentle scrubbing action, cleaning the interior surfaces of the colorimeter
cell. Clean the stir bar with the cotton swab (
Figure 23
).
9.
Install the stir bar.
10.
Replace the rubber plug on the top of the colorimeter assembly and make sure the
cover is on the viewing port of the colorimeter.
11.
Press the
EXIT
key to immediately return to normal operation; the instrument will
return to normal operation automatically after 60 minutes.
*
Do not use a plastic swab when cleaning the colorimeter with sulfuric acid. The the acid will dissolve the plastic.
Summary of Contents for Lange CL17
Page 2: ......
Page 13: ...11 Installation Figure 3 CL17 instrument dimensions 2 of 3 ...
Page 18: ...16 Installation Figure 7 Sample conditioning kit ...
Page 32: ...30 System Startup ...
Page 40: ...38 Operation ...
Page 50: ...48 Troubleshooting ...
Page 52: ...50 Parts and accessories ...
Page 56: ...54 Certification ...
Page 60: ...58 Network Interface Card ...