24
Installation
5.
Remove the connector from the instrument. Refer to
Figure 16 on page 24
for
connector position.
6.
Insert the wire ends into the connector (refer to the table below) until the insulation
seats against the connector as shown in
Figure 14
. (Do not seat the insulation under
the terminal pinch plates.)
7.
Replace the connector and supply power to the instrument.
3.9 Install the pump/valve pinch plate
To eliminate the effects of constant pressure on the pump tubes during shipment and
storage, the pinch plate and screws for the pump/valve module are shipped in the
installation kit and the pump tubes are held in place with tape.
Reagent flow through the pump/valve module must be from bottom to top. If reversed,
fluid mixture will be pumped from the colorimeter sample cell, causing reagent bottles to
overflow.
Complete the assembly of the pump/valve module as follows:
1.
Remove the tape.
2.
Make sure the individual pinch blocks are positioned as shown in
Figure 17
, with the
dissimilar-shaped block on top.
3.
Align the pinch plate on the pump/valve module (refer to
Figure 18
).
Recorder Wires
Circuit Board Markings
Re
+
Recorder –
–
Shield
GND
Figure 16 Recorder connections
1
Make sure no power is supplied to the instrument
2
Make sure no power is supplied to the relay
Summary of Contents for Lange CL17
Page 2: ......
Page 13: ...11 Installation Figure 3 CL17 instrument dimensions 2 of 3 ...
Page 18: ...16 Installation Figure 7 Sample conditioning kit ...
Page 32: ...30 System Startup ...
Page 40: ...38 Operation ...
Page 50: ...48 Troubleshooting ...
Page 52: ...50 Parts and accessories ...
Page 56: ...54 Certification ...
Page 60: ...58 Network Interface Card ...