41
Maintenance
6.2
Unscheduled maintenance
6.2.1
Fuse replacement
The T, 2.5A, 250V fuse used in this instrument is used for both 115V and 230V operation.
DANGER
Electrocution hazard. Remove power from the instrument when removing or
installing a fuse.
DANGER
Fire hazard. For continued protection against fire, replace the fuse only with a fuse
of the same type and rating.
Fuse replacement procedure:
1.
Make sure there is no power supplied to the instrument. the instrument power switch
does not remove power from the fuses, Power must be disconnected remotely to
remove power from the fuses.
2.
Remove the customer access cover.
3.
Locate the fuse holders (near the terminal strip in the customer wiring compartment).
Refer to
Figure 13 on page 22
.
4.
Remove the two fuses (F1 and F2) and replace them with two new fuses with the
same specifications (T, 2.5A, 250V). Refer to
Parts and accessories on page 49
.
5.
Reinstall the customer access cover and resupply power.
Table 7 Replacement Tubing Lengths for
Figure 21
Item
Description
Length (Qty)
From...
To...
Item Number
1
1/8"
ID, 1/4" OD
4.5 inches (1)
Pump Body Out
Colorimeter
43293-00
2
1/32" ID, 3/32" OD
7.0 inches (2)
Pump Body Out
Y-Fitting
44253-00
3
1/16" ID, 3/32" OD
2.0 inches (4)
Pump Body In
Pump Body Out
42717-00
4
0.062 ID, 0.125" OD
6.0 inches (2)
Reagent Bottle Cap
Pump Body In
42076-00
5
0.062" ID, 0.125" OD
6.0 inches (2)
Reagent Bottle Cap
Reagent Vent Fitting
42076-00
6
1/32" ID, 3/32" OD
7.0 inches (2)
Reagent Bottle Bottom
Reagent Bottle Cap
45524-00
7
1/8" ID, 1/4" OD
7.0 inches (1)
Sample Bypass Tee
Pump Body In
43293-00
8
1/8" ID, 1/4" OD
1.5 inches
Sample Bypass Tee
Sample Inlet Fitting
43293-00
9
1/4" OD x 0.04 W, Black
varies (1)
Sample Conditioning Out
Case Fitting
30616-00
10
1/2"
ID
varies (1)
Instrument Drain
Customer Drain
(not supplied)
11
1/32"
ID, 3/32"
OD
3 inches
Sample Bypass Tee
Drain fitting
44253-00
12
1/32" ID, 3/32" OD
1.0 inch (1)
Y-Fitting
Colorimeter
44253-00
13
0.500” ID, 11/16” OD
12 inches
Colorimeter
Drain Fitting
54108-00
Table 8 Drain, air purge, sample-in, and enclosure locations
A
Air purge, 0.01 CFM at 20 PSIG max, ¼ in. O.D. tube
B
Case drain, ½ in. I.D. tube
C
Sample drain, ½ in. I.D. tube
D
Sample in, 5 PSIG max, ¼ in. O.D. tube
E
Enclosure
Summary of Contents for Lange CL17
Page 2: ......
Page 13: ...11 Installation Figure 3 CL17 instrument dimensions 2 of 3 ...
Page 18: ...16 Installation Figure 7 Sample conditioning kit ...
Page 32: ...30 System Startup ...
Page 40: ...38 Operation ...
Page 50: ...48 Troubleshooting ...
Page 52: ...50 Parts and accessories ...
Page 56: ...54 Certification ...
Page 60: ...58 Network Interface Card ...