Specification
Details
Resolution
Turbidity: 0.001 NTU/EBC
Nephelo: 0.01
Absorbance: 0.001 Abs
Transmittance: 0.1% T
Color: 1 CU
Repeatability
±1% of reading or 0.01 NTU, whichever is greater
(under reference conditions)
Response time
Signal averaging off: 6.8 seconds
Signal averaging on: 14 seconds (when
10 measurements are used to calculate the average)
Stabilization time
Ratio on: 30 minutes after start-up
Ratio off: 60 minutes after start-up
Reading modes
Manual or auto range, signal averaging on and
adjustable or off, Ratio on or off
Power requirement
115–230 VAC, 50/60 Hz (automatic power selection)
28 W maximum
Pollution
degree/installation
category
2; II
Protection Class
1
Operating conditions
Temperature: 0 to 40 °C (32 to 104 °F)
Relative humidity: 0–90% at 25 °C, 0–75% at 40 °C,
noncondensing
Altitude: 2000 m (6560 ft) maximum
Indoor use only
Storage conditions
–40 to 60 °C (–40 to 140 °F), instrument only
Printer
Built-in (thermal, 58-mm, up to 28 column)
Interface
RS232C serial interface by way of DB9 subminiature
D-shell connector for data output to computer or
printer, and data input (command). No handshaking.
Specification
Details
Air purge
Dry nitrogen or instrument grade air (ANSI MC 11.1,
1975)
0.1 scfm at 69 kPa (10 psig); 138 kPa (20 psig)
maximum
Hose barb connection for
1
/
8
-inch tubing
Sample cells
Round cells 95 x 25 mm (3.74 x 1 in.) borosilicate
glass with rubber-lined screw caps
Note: Smaller sample cells (less than 25 mm) can be used
when a cell adapter is used.
Sample requirements
25 mm sample cell: 20 mL minimum
0 to 95 °C (32 to 203 °F)
Enclosure
High-impact polycarbonate plastic
Dimensions
30.5 x 40 x 15.6 cm (12.0 x 15.7 x 6.1 in.)
Weight
3.8 kg (8.5 lb)
Certification
CE, cETLus
General information
In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without
notice or obligation. Revised editions are found on the manufacturer’s
website.
Additional information
Additional information is available on the manufacturer's website.
4
English
Summary of Contents for Lange 2100AN
Page 18: ...1 18 English...
Page 39: ...1 Deutsch 39...
Page 60: ...1 60 Italiano...
Page 80: ...1 80 Fran ais...
Page 102: ...1 102 Espa ol...
Page 122: ...1 122 e tina...
Page 142: ...1 142 Dansk...
Page 162: ...1 162 Nederlands...
Page 182: ...1 182 Polski...
Page 202: ...1 202 Svenska...
Page 222: ...1 222 Suomi...
Page 241: ...1 Magyar 241...
Page 261: ...1 Rom n 261...
Page 281: ...1 T rk e 281...
Page 288: ...12 2005 2002 96 IECS 003 A IECS IECS FCC 15 A 15 FCC 1 2 15 FCC 1 2 3 4 5 288...
Page 292: ...1 Calibration Setup 1 NTU NEP EBC CU T Abs 292...
Page 296: ...StablCal 1 302 2 USEPA 302 3 4 CAL Zero CAL Zero 00 NTU 5 0 1 NTU 6 296...
Page 298: ...30 60 24 298...
Page 299: ...1 30 mL 2 3 4 5 6 299...
Page 300: ...7 RS232 PRINT RANGE MANUAL AUTO 40 NTU 268 nephelo 9 8 EBC 300...
Page 302: ...1 2 3 1 2 3 302 4 5 1 2 3 30 60 302...
Page 303: ...1 303...
Page 307: ......