Figure 15 Total report for sample data—3 lines
1
Method name for probe settings
4
Calibration slope/ratio/constant
2
Time of calibration, prefaced by “CAL” and displayed as hh:mm in 24 h (or user-
defined) format
5
Offset—contents vary depending on type of parameter and user settings. May
be blank.
3
Date of calibration (DD-MM-YY or user-defined format)
6
Operator ID: user-defined; shows “- - -” if undefined
Check standard reports
Check standard data is printed with 1 line of information (
Figure 16
)
Figure 16 Check standard report
1
Report type (ST = check standard)
5
Measured value
2
Report type (ST = check standard)
6
Units
3
Time (hh:mm:ss in 24 h or user-defined format)
7
Check standard status: Pass/Fail based on the acceptance criteria
4
Date (DD-MM-YY or user-defined format)
8
Operator ID: user-defined; shows “- - -” if undefined
Calibration reports
Calibration data is printed when the data log is sent to the printer or when probe data is sent to the printer. Calibration data is printed with 2 lines of
information (
Figure 17
).