Operation
20
4.4 ERROR CODES
Some error codes, such as poor electrode condition, can be overridden by the user
to allow the instrument to be used until a replacement electrode can be obtained.
When an error is displayed, first check to determine the cause of the error. To abort
the calibration or other error causing condition, Push
SETUP
or the
pH
COND/TDS/SAL) key.
To override an error code (not recommended):
1.
Push
ENTER.
2.
Push the up
arrow for display to
YES.
This indicates the option to perform an
error override. Push
ENTER
. Refer to
section Section 8 on page 41
for
troubleshooting error codes, causes and actions.
N O T I C E
It is not recommended that the user override the error codes. It is not possible to override all
error codes. Pushing the arrow keys will display the
Yes
and
NO
options.