Maintenance
39
7.4 Probe cleaning
7.4.1 ISFET non-glass probe cleaning and reconditioning
7.4.1.1
Cleaning the ISFET non-glass probes
1.
Regularly clean the ISFET probe with detergent (a few drops of dish detergent
in a cup of warm water).
2.
Rinse well with deionized water.
N O T I C E
Do not use metal objects such as pins or paperclips to clean the sensor. The sensor can be
permanently damaged with aggressive abrasion of the sensor surface.
Figure 5 Probe cleaning
N O T I C E
Do not use acetone, toluene, or other strong organic solvents on ISFET probes.