Error
Description
User action recommendations
Technical information including troubleshooting action
items (only for Service Operator)
E-10
Environment
temperature too
high or too low
The ambient temperature is out
of range. Use the instrument only
in the specified operating
conditions. Refer to
on page 3.
The ambient temperature was measured out of range. Probably
the instrument was used out of spec temperature. (T < -20°C or
T > 70 °C)
+ Check/read the service log for possible other error messages.
>
E-12
Low battery
power
Battery power is too low.
Replace the batteries. Refer to
Battery power is too low. Error appears between 0-2% battery
capacity.
+ Check/read the service log for possible other error messages.
Check battery power.
> Replace batteries
Check battery contacts got appropriate contact to mainboard.
>
on page 49
>
E-13
Parameter load
failure
The memory of the instrument is
defective. Contact technical
support.
The DR300 chemical parameters could not be loaded (neither
from filesystem nor from backup). The device is not capable to
function without and is therefore locked.
+ Check/read the service log for possible other error messages.
> Update mainboard firmware and parameter setting.
>
E-14
Zero
measurement
invalid
The zero measurement is too
low. Use a sample cell filled with
water and try again. If the error
continues, contact technical
support.
Signal of zero measurement is too low. Zero - High Frequency
< 1 KHz
+ Check/read the service log for possible other error messages.
> Use water vial and try again.
> Replace LED
> Replace optical filter
> Replace detector
>
E-15
Absorbance too
high
Identify if there is a light blockage
in the cell holder. Clean the cell
holder. Dilute the sample. Do the
test again.
Note: This instrument can not read
absorbance values higher than
3.5 Abs.
Absorbance was read > 3.5 Abs.
+ Check/read the service log for possible other error messages.
Check if beam path (cell compartment) is blocked.
> Clean cell compartment.
Check the absorbance reading in 'service mode'.
Check absorbance values for service test kit vials.
> Replace detector
>
> Replace LED
> Replace optical filter
E-20
Signal
measurement
out of range
There is too much light on the
light detector. Make sure that the
instrument cap is fully installed
on the cell holder. Do the test
again. If the error continues,
contact technical support.
Too much signal on light detector. Zero or Read signal - High
frequency > 500 KHz
+ Check/read the service log for possible other error messages.
Check the absorbance reading in 'service mode'.
Check absorbance values for service test kit vials.
Check optical aperture in front of LED is still in place.
> Replace optical aperture
> Replace optical filter
> Replace LED
> Replace detector
>
Troubleshooting
55
Summary of Contents for DR300
Page 1: ...DOC222 53 90639 DR300 Service Manual 02 2019 Edition 2...
Page 2: ......
Page 4: ...Table of Contents 2...
Page 6: ...Specifications 4...
Page 15: ...Startup 13...
Page 16: ...Startup 14...
Page 36: ...Inspection 34...
Page 54: ...Repairs 52...
Page 65: ...Figure 15 Set delivery state Troubleshooting 63...
Page 66: ...Troubleshooting 64...
Page 82: ...Certificate of Service Inspection 80...
Page 84: ...Certificate of Service Inspection manually 82...
Page 85: ......