Step Description
Picture
4
Insert the UART connector of the cable.
5
Push the connector down.
6
Connect the USB connector of the cable to a USB-A port of
the PC.
7
Start the service tool software VAP235.
4.5.3 Do an inspection with the test filter set, service cable, PC and VAP235 software
Items to collect:
•
PC and Software VAP235
•
Service cable VAA10036.99.G00
•
Test filter set service DR300 VAA10038.99.V00
•
Printer (if a printout of the Certificate of Service Inspection is necessary)
Inspection
21
Summary of Contents for DR300
Page 1: ...DOC222 53 90639 DR300 Service Manual 02 2019 Edition 2...
Page 2: ......
Page 4: ...Table of Contents 2...
Page 6: ...Specifications 4...
Page 15: ...Startup 13...
Page 16: ...Startup 14...
Page 36: ...Inspection 34...
Page 54: ...Repairs 52...
Page 65: ...Figure 15 Set delivery state Troubleshooting 63...
Page 66: ...Troubleshooting 64...
Page 82: ...Certificate of Service Inspection 80...
Page 84: ...Certificate of Service Inspection manually 82...
Page 85: ......