Step Description
Picture
17
The measured value of vial 8 (in absorbance) is shown.
Service inspection finished.
18
If the inspection completes without error, a Certificate of
Service Inspection HDQ303 is automatically stored. Find the
pdf file in the path which is defined in Settings > Program
Settings > General > Certificate path. Print it out and sign.
Refer to
Certificate of Service Inspection
on page 79.
Subsequently, an instrument log is automatically read out.
Refer to
on page 57. Find the txt file in the path
which is defined in Settings > Program Settings > General >
Instrument log path. E.g. Logger_LPG445_18060A000047.txt.
This inspection task is also available as a service video on HSO. HSO full access
requires login credentials. First log into HSO and afterwards click on
https://support.hach.com/app/answers/answer_view/a_id/1024331
. Select the desired
service video.
Note: Following hyperlink without being logged will end in a permission denied information
screen.
4.6 Do a service inspection with the test filter set and service mode
Items to collect:
•
VAA10038.99.V00 Test filter set Service DR300
•
Certificate of Service Inspection, refer to
Certificate of Service Inspection (manually)
on page 81
Step Description
Picture
1
Note the part number and serial number from the
DR300 on the Certificate of Service Inspection.
2
Note the set number and expiration date from
VAA10038.99.V00 on the Certificate of Service
Inspection.
Inspection
28
Summary of Contents for DR300
Page 1: ...DOC222 53 90639 DR300 Service Manual 02 2019 Edition 2...
Page 2: ......
Page 4: ...Table of Contents 2...
Page 6: ...Specifications 4...
Page 15: ...Startup 13...
Page 16: ...Startup 14...
Page 36: ...Inspection 34...
Page 54: ...Repairs 52...
Page 65: ...Figure 15 Set delivery state Troubleshooting 63...
Page 66: ...Troubleshooting 64...
Page 82: ...Certificate of Service Inspection 80...
Page 84: ...Certificate of Service Inspection manually 82...
Page 85: ......