Figure 3 AC power connections
Table 3 AC wiring information
Pin
Connection
Color—North America
Color—EU, UK, AU
1
Protective earth ground (G)
Green
Green with yellow stripe
2
Neutral (N)
White
Blue
3
Hot (L)
Black
Brown
Connect to DC power
Use a minimum 18 AWG wire to connect to DC power from a solar panel or customer-supplied DC
power.
1.
Open the controller cover. Refer to
2.
Put the power cable through a strain relief fitting near the DC power connector. Refer to
.
3.
Tighten the strain relief fitting.
4.
Close the controller cover.