ATTENTION!
The tolerances / measurement precision specified here only apply for use of the devices in EMC en-
vironments in compliance with the basic requirements in accordance with DIN EN 61326-1:2013.
Only use the supplied mains adapter!
Use of an incorrect mains adapter can result in damage to the measurement system and/or the stirrer
drive.
Unauthorised modifications of the system can result in measurement errors, danger, and electroma-
gnetic interference. The manufacturer bears no responsibility for the consequences of such modifica-
tions.
Do not place hot stirring containers on the stirrer drive. Maximum temperature: 56°C. The stirrer drive
can be damaged if this is disregarded.
The device may only be opened by an authorised service location in the case of a repair. Disconnect
from the mains before opening!
The test bottles are contained in the measurement system during transport. Make sure that they do
not fall to the floor during unpacking.
IMPORTANT INFORMATION!
Do not leave a magnetic stir bar in an alternating magnetic field when it cannot rotate. It can be
exposed to a powerful opposing magnetic field. It could be demagnetised as a result.
Observe the permissible environmental conditions as specified in the technical data.
Avoid extreme temperature changes.
Do not use the mains adapter in damp rooms and do not use it in areas subject to spray water.
If the mains adapter has become wet or damp, do not touch it or the system. De-energise the socket.
Only use the mains adapter for the power supply of a stirrer drive and a measuring unit. If more de-
vices are connected, the system can malfunction.
The length of the magnetic stir bars should not exceed 40 mm. Do not use elliptical stir bars with a
round cross-section.
Place the stirrer drive on a horizontal surface with sufficient bearing capacity to support the weight of
the stirrer drive and the measurement system with filled bottles. The measurement system can slide
off the stirrer platform if it is in an inclined position. Bottles and the system can fall to the floor.
To disconnect the mains adapter, only pull out the secondary cable (15V DC) from the measurement
system and and the stirrer drive at the plugs! Otherwise, the cable can be damaged.
Remove the depleted batteries from the device soon in order to avoid leakage.
For periods of extended non-use, remove the batteries from the device as a precautionary measure.
The battery status should be checked regularly, even if the device is normally supplied externally
(mains adapter plug), in order to prevent leaks from empty batteries.
The sample may not come into contact with KOH. Otherwise the measurements will be falsified.
The stirrer drive starts automatically as soon as the power supply is connected. Any magnetic stir bars
or other magnetic parts nearby can move unexpectedly as a result.
GB
Summary of Contents for BOD Direct plus
Page 1: ...DOC022 98 90554 BOD System BOD Direct plus 03 2016 Edition 2 Basic User Manual...
Page 146: ...146 BOD Direct plus_BMU_2 03 2016 2012 19 E Hach 13 08 2005 Hach 2006 66 E BG...
Page 147: ...147 BOD Direct plus_BMU_2 03 2016 KOH BG...
Page 148: ...148 BOD Direct plus_BMU_2 03 2016 MSDS SDB EM BSB BG...
Page 149: ...149 BOD Direct plus_BMU_2 03 2016 1 EM DIN EN 61326 1 2013 56 C 40 15V DC KOH BG...
Page 150: ...150 BOD Direct plus_BMU_2 03 2016 2 3 3 1 BSB5 BSB7 OECD 301F 1 28 3 2 1 BG...
Page 152: ...152 BOD Direct plus_BMU_2 03 2016 5 5 1 1 2 3 4 1 2 3 5 6 ON OFF 7 BG...
Page 154: ...154 BOD Direct plus_BMU_2 03 2016 5 2 3 C LR14 BG...
Page 157: ...157 BOD Direct plus_BMU_2 03 2016 6 BSB BSB BSB 4 BSB BSB BSB BSB BG...
Page 158: ...158 BOD Direct plus_BMU_2 03 2016 7 7 1 BSB 6 40 7 2 DC Y Y Y BG...
Page 159: ...159 BOD Direct plus_BMU_2 03 2016 7 3 56 C BSB BSB 7 4 PVC 8 BG...
Page 162: ...162 BOD Direct plus_BMU_2 03 2016 2012 19 Hach 13 08 2005 Hach 2006 66 GR...
Page 163: ...163 BOD Direct plus_BMU_2 03 2016 KOH GR...
Page 164: ...164 BOD Direct plus_BMU_2 03 2016 MSDS SDB BOD GR...
Page 165: ...165 BOD Direct plus_BMU_2 03 2016 1 DIN EN 61326 1 2013 56 C 40 mm 15V DC KOH GR...
Page 166: ...166 BOD Direct plus_BMU_2 03 2016 2 3 3 1 BOD5 BOD7 OECD 301F 1 28 3 2 1 GR...
Page 170: ...170 BOD Direct plus_BMU_2 03 2016 5 2 3 C LR14 Baby GR...
Page 173: ...173 BOD Direct plus_BMU_2 03 2016 6 BOD BOD BOD 4 BOD BOD BOD BOD GR...
Page 174: ...174 BOD Direct plus_BMU_2 03 2016 7 7 1 BOD 6 40 7 2 DC Y Y Y GR...
Page 175: ...175 BOD Direct plus_BMU_2 03 2016 7 3 56 C BOD BOD 7 4 PVC 8 GR...