18. Cal Signal (if configured in the system).
Use this function to test the Calibration Signal output. To
change the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER
key again. If the device is on, it will be marked with an “*”. See Section
19. Maint Signal (if configured in the system).
Use this function to test the Maintenance Signal output. To
change the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER
key again. If the device is on, it will be marked with an “*”. See Section
20. Remote Standby (if configured in the system).
Use this function to test the Remote Standby output. To
change the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER
key again. If the device is on, it will be marked with an “*”. See Section
21. Stop (if configured in the system).
Use this function to test the Stop output. To change the state of the
device, press the ENTER key, set the device to ON/OFF, and press the ENTER key again. If the device is
on, it will be marked with an “*”. See Section
22. Note (if configured in the system).
Use this function to test the system Notification output. To change
the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER key
again. If the device is on, it will be marked with
23. Warning (if configured in the system).
Use these functions to test the Warning output. To change the
state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER key again. If
the device is on, it will be marked with an “*”. See Section
24. Fault (if configured in the system).
Use this function to test the Fault output. To change the state of the
device, press the ENTER key, set the device to ON/OFF, and press the ENTER key again. If the device is
on, it will be marked with an “*”. See Section
25. Man Mode Trig (if configured in the system).
Use this function to test system Manual Mode Trigger
output. To change the state of the device, press the ENTER key, set the device to ON/OFF, and press
the ENTER key again. If the device is on,
it will be marked with an “*”. See Section
26. 4-20mA Chng (if configured in the system).
Use this function to test system generic 4-20mA Change
output signal. To change the state of the device, press the ENTER key, set the device to ON/OFF, and
press the ENTER key again. If the device is on, it will be marked with an “*”. See Section
27-29. 4-20mA Chng 1-3 (if configured in the system).
Use this function to test system 4-20mA Change
output signal for stream specific channels from Channel 1 to Channel 3. To change the state of the
device, press the ENTER key, set the device to ON/OFF, and press the ENTER key again. If the device is
on, it will be marked with an “*”. See Section
30. 4-20mA Read (if configured in the system).
Use this function to test system 4-20mA Read output. To
change the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER
key again. If the device is on, it will be marked with an “*”.See Section
32.
Input/Output Status.
Input/Output Status is a link to Maintenance, Diagnostics, Input/Output Status
Summary of Contents for BioTector B3500ul
Page 17: ...Page 17 Software Menu Diagram...
Page 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Page 63: ...Page 63...
Page 75: ...Page 75 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 157: ...Page 157 Section 11 System Replacement and Spare Parts...
Page 166: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...