Preparation of Calibration Standard Solutions:
Use eye protection and gloves.
Standards solutions greater than 1000mg/l can be prepared directly without any dilution by simply mixing the
necessary quantity solvent or salt with deionized water. Standard solutions below 1000mg/l concentration
should be prepared by dilution technique. First a 1000mg/l standard stock solution should be prepared, and
then the required lower concentration standard solution should be prepared by applying the necessary dilution
procedures:
➢
For example, to prepare a 50mgC/l TOC only standard solution, first weigh 50 grams of the
1000mgC/l stock standard. Add 50 grams of the 1000mgC/l standard into a one-liter flask and add
enough deionized water to make the solution exactly 1 liter.
➢
For increased accuracy, standard solutions below 5mg/l (ppm) concentration should be prepared with
two or more steps dilution. For example, to prepare a 1mgC/l standard, first prepare a 100mgC/l
standard by adding 100 grams of the 1000mgC/l standard into a one-liter flask and by adding enough
deionized water to make it exactly a 1 liter solution. Then add 10 grams of the 100mgC/l standard into
a one-liter flask and add enough deionized water to make it exactly 1 liter.
➢
Standard solution
s at μg/l (ppb) levels should be prepared with several dilution steps. For instance, a
1mgC/l (1000 μg/l) standard should be prepared with two or more steps dilution as described above.
To prepare a 50μg/l standard, add 50 grams of the 1000μg/l standard into a one-liter flask, and add
enough deionized water to make it exactly 1 liter.
Shelf Life and Storage of Calibration Standard Solutions:
❖
TOC standards prepared from Potassium Hydrogen Phthalate is typically stable for a month once it
is kept in a closed glass container and refrigerated at 4°C.
❖
All other standards such as TOC prepared with Acetic Acid and TIC standards are recommended to
be used within 48 hours of manufacture.
Summary of Contents for BioTector B3500ul
Page 17: ...Page 17 Software Menu Diagram...
Page 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Page 63: ...Page 63...
Page 75: ...Page 75 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 157: ...Page 157 Section 11 System Replacement and Spare Parts...
Page 166: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...