![Hach BioTector B3500ul User Manual Download Page 132](http://html1.mh-extra.com/html/hach/biotector-b3500ul/biotector-b3500ul_user-manual_3204050132.webp)
Page 132
8.3.10
CO
2
Analyzer
ANALYSIS GRAPH SCALE 2000ppm
▪
ANALYSIS GRAPH SCALE (2000ppm by default) determines
the scale of the y-
axis “CO2 analyzer’s ppm
-
CO2 readings” on
the Analysis Graph Screen (see section
for details). This function allows the system to display
CO2 data peaks in optimal resolution on the LCD screen.
▪
The ANALYSIS GRAPH SCALE is independent of the CO2
ANALYZER RANGE described below. Note that, in certain
circumstances, when the displayed CO2 peaks on screen
exceeds the scale of the graph, the CO2 analyzer keeps
measuring and integrating the CO2 readings to obtain the TOC
result, without losing any CO2 data.
BAUDRATE 9600
▪
BAUDRATE (9600bps by default) specifies the data
communication signaling speed of CO
2
Analyzer with the RS232
communication interface.
CO2 ANALYZER RANGE 1000ppm
▪
CO2 ANALYZER RANGE defines the full scale range of the CO
2
analyzer installed in the BioTector.
CO2 ANALYZER CAL
▪
The CO2 ANALYZER CALIBRATION menu allows the user to
change CO
2
Analyzer range and the CO
2
analyzer zero and
span parameters when necessary. If required, please contact
manufacturer or distributor for details of the procedures.
Summary of Contents for BioTector B3500ul
Page 17: ...Page 17 Software Menu Diagram...
Page 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Page 63: ...Page 63...
Page 75: ...Page 75 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 157: ...Page 157 Section 11 System Replacement and Spare Parts...
Page 166: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...