8.2.5 New Reagents Program
SPAN CALIBRATION
NO
SPAN CALIBRATION (NO by default) defines if the Span
Calibration reactions are part of the Install New Reagents
procedure in Reagents Setup menu.
If SPAN CALIBRATION is activated (YES), the standard solution
needs to be connected to the Calibration/Manual port of the
BioTector as part of the Install New Reagents procedure. See
concentrations of the TOC STANDARD defined in Span
Calibration menu (see Section
for details).
SPAN CHECK NO
SPAN CHECK (NO by default) defines if the Span Check
reactions are part of the Install New Reagents procedure in
Reagents Setup menu.
Note that both SPAN CALIBRATION and SPAN CHECK
functions cannot be selected as YES simultaneously and cannot
occur together in the Install New Reagents procedure.
AUTOMATIC RE-START YES
AUTOMATIC RE-START (Yes by default) defines BioTector to
return automatically online operation having completed the
Install New Reagents cycle.
8.2.6 Reagents Monitor
REAGENTS MONITOR
YES
When REAGENTS MONITOR is activated (YES), the system
creates a new Reagent Status screen which displays the amount
of reagents and the estimated number of days left for each
reagent to last. The Reagent Status screen is displayed for a
period of 15 minutes and the system automatically reverts back
to the default Reaction Result screen.
BioTector calculates the reagent usage and the days remaining
based on the average reagent usage from the reaction archive.
If there are no reactions available in the archive, BioTector
calculates the reagent usage using the operation range
information in the system program menus and updates the
reagent status figures accordingly.
Note that when one or more reagent volumes are updated in this
menu, system automatically resets the new reagents volumes in
Install New Reagents menu and also updates the figures
displayed in the main Reagents Status screen.
LOW REAGENTS NOTE
LOW REAGENTS, which is programmed as NOTE (notification)
by default, will activate a relay specifically programmed for
notification events.
When LOW REAGENTS is programmed as WARNING, and if
reagents run low, system generates an “85_LOW REAGENTS”
warning and activates a relay specifically programmed for
warning events.
LOW REAGENTS AT 20 DAYS
LOW REAGENTS AT specifies the number of days (20 days by
default), before the system generates an “85_LOW REAGENTS”
warning or notification, depending on the LOW REAGENT
setting above.
This function can be used by the user as an early warning or
notification of system low reagents so that the required reagents
can be prepared or ordered in time.
NO REAGENTS WARNING
NO REAGENTS programmed as a WARNING event by default,
can also be set as a FAULT event. When system calculates no
reagents, the fault relay is activated and system generates a
“20_NO REAGENTS” fault or warning.
NO REAGENTS can also be programmed as NOTE
(notification) where only a relay programmed as notification
operates in the event of no reagents.
Summary of Contents for BioTector B3500C
Page 17: ...Page 17 Software Menu Diagram...
Page 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Page 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Page 63: ...Page 63...
Page 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Page 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Page 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...