Figure 9 Site installation
1
Strainer
2
Vertical lift
3
Mounting surface
3.4 Electrical installation
3.4.1 Connect the sampler to power
D A N G E R
Electrocution hazard. If this equipment is used outdoors or in potentially wet locations, a
Ground Fault Circuit Interrupt (GFCI/GFI) device must be used for connecting the
equipment to its main power source.
D A N G E R
Fire hazard. Install a 15 A circuit breaker in the power line. A circuit breaker can be the
local power disconnect, if located in close proximity to the equipment.
D A N G E R
Electrocution hazard. Protective Earth Ground (PE) connection is required.
W A R N I N G
Electrocution hazard. Make sure that there is easy access to the local power disconnect.
Connect the power cord on the AWRS or both power cords on the refrigerated sampler. The
refrigerator starts after a 5-minute delay. Use a power line filter or connect the power cord for the
controller to a different branch circuit to decrease the possibility of electrical transients.