Figure 6 Multiple bottle installation
1
Retainer for 24 1-L poly
bottles
4
Bottle tray for 8 to 24 bottles
7
Front of sampler
2
Retainer for 24 350-mL
glass bottles
5
Slot for bottle tray
8
Removable tray (AWRS
only)
3
Bottle one indicator
6
Floor of refrigerated
sampler
9
Insert (refrigerated sampler
only)
3.3 Plumb the sampler
Install the intake tube in the middle of the sample stream (not near the surface or bottom) to make
sure that a representative sample is collected.
1.
For a sampler with the standard liquid detector, connect the tubing to the sampler as shown in
Note: When Teflon-lined tubing is used, use the tubing connection kit for Teflon-lined PE tubing.
2.
For a sampler with the optional non-contacting liquid detector, connect the tubing to the sampler
as shown in
Note: When Teflon-lined tubing is used, use the tubing connection kit for Teflon-lined PE tubing.
3.
Install the intake tube and strainer in the main stream of the sample source where the water is
turbulent and well-mixed. Refer to
• Make the intake tube as short as possible. Refer to
intake tubing length.