Table 1 General specifications (continued)
Specification
Details
Air purge
15 scfh (7 L/min), dry instrument air, inert gas
Certifications
Certified to UL and CSA by ETL, CE
1
Six 1-liter bottles of water = 6 kg. The small amount of fluids in the analyzer =
0.5 kg. Add another 0.5 kg for chemicals that weigh more than water.
Table 2 Measurement specifications (part 1 of 2)
Specification
Alkalinity
Ammonia LR
Ammonia/
Monochloramine
Light source
Class 1M LED,
peak wavelength
600 nm
Class 1M LED,
peak wavelength
650 nm
Class 1M LED,
peak wavelength
650 nm
Parameters
measured
Total (T) alkalinity
Phenolphthalein
(P) alkalinity
Hydroxide
alkalinity
Carbonate
alkalinity
Bicarbonate
alkalinity
Free ammonia
1
Total ammonia,
monochloramine
Measurement
range
T alkalinity:
1–1000 mg/L as
CaCO
3
P alkalinity:
5–1000 mg/L as
CaCO
3
0.02–2 mg/L as N
0.1–10 mg/L as
Cl
2
Accuracy
2
±5% or ±1.0 mg/L
(the larger value)
±5% or
±0.02 mg/L (the
larger value)
±5% or
±0.10 mg/L (the
larger value)
Precision/
Repeatability
±3% or ±0.6 mg/L
(the larger value)
±3% or
±0.01 mg/L (the
larger value)
±3% or
±0.05 mg/L (the
larger value)
Table 2 Measurement specifications (part 1 of 2) (continued)
Specification
Alkalinity
Ammonia LR
Ammonia/
Monochloramine
Response time
< 10 minutes
3
< 5 minutes
3
< 5 minutes
3
Cycle time
(average)
8 minutes
6.2 minutes
6.2 minutes
Minimum
detection limit
≤ 0.10 mg/L
≤ 0.01 mg/L
≤ 0.05 mg/L
1
The ammonia analyzer measures background monochloramine during
calibration only.
2
Based on a calibration frequency of once a day
3
For a 90% response to a step change at the sample inlet and one sample
stream
Table 3 Measurement specifications (part 2 of 2)
Specification
Hardness LR
Hardness HR
Light source
Class 1M LED, peak
wavelength 520 nm
Class 1M LED, peak
wavelength 600 nm
Parameters measured
Total hardness
Total hardness
Measurement range
0.05–10 mg/L as CaCO
3
10–1000 mg/L as CaCO
3
Accuracy
1
±5% or ±0.05 mg/L (the
larger value)
±5% or ±2 mg/L (the larger
value)
Precision/ Repeatability ±3% or ±0.03 mg/L (the
larger value)
±5% or ±2 mg/L (the larger
value)
Response time
< 5 minutes
2
< 17 minutes
2
Cycle time (average)
4 minutes
8.2 minutes
Minimum detection limit 0.05 mg/L
10 mg/L
1
Based on a calibration frequency of once a day
2
For a 90% response to a step change at the sample inlet and one sample
stream
4
English
Summary of Contents for APA 6000
Page 1: ...DOC023 97 80365 APA 6000 05 2013 Edition 1 Installation Installation Instalaci n...
Page 2: ...English 3 Fran ais 21 Espa ol 40 2...
Page 30: ...Figure 9 Montage sur une table Figure 10 Montage sur un mur 30 Fran ais...
Page 49: ...Figura 9 Montaje de sobremesa Figura 10 Montaje en pared Espa ol 49...
Page 60: ......
Page 61: ......