30
English
Troubleshooting
7.3 Troubleshooting the Electrode
The electrode data are saved in the AMTAX sc Calibration data menu or in the event
logger.
Typical electrode values (note sign):
contains typical electrode data for a new inserted electrode with new electrolyte
and membrane cap.
The zero value is always the most positive value and the standard 2 value is always the
most negative value in a calibration. The standard 1 is always between the zero value and
the standard 2 value.
When starting up an electrode, the slope increases to the final value (ideal between –58
and –63 mV) and remains there with minor fluctuations.
The error descriptions for the electrode in
and
are related to the
error/warning "electrode slope" or "false elec. data".
Table 7Typical electrode values
Description
MR 1
(0.02–5 mg/L NH4–N)
MR 2
(0.05–20 mg/L NH4–N)
MR 3
(1–100 mg/L NH4–N)
MR 4
(10–1000 mg/L
NH4–N)
Slope
–55 to –67 mV
Slope warning
–50 to –55 mV or –67 to –70 mV
Slope error
0 to –50 mV or –70 to –150 mV
mV zero
–205 to +5 mV
20 to 120 mV
20 to 120 mV
20 to 120 mV
mV standard 1
–200 to –140 mV
–30 to 30 mV
–25 to –85 mV
–70 to –130 mV
mV standard 2
–240 to –180 mV
–30 to –90 mV
–70 to –130 mV
–130 to –190 mV
Table 8Error messages
Error description
Diagnostics
Solution
The calibration values for the standards
are both above the mV zero value.
- Reagent empty
- Reagent pump faulty
- Fittings leaking
- New reagent
- Replace reagent pump
- Tighten fittings
The calibration values for the two
standards and the mV zero value are all
indicating very similar values.
- Electrode faulty
- Electrolyte empty
- Re-new electrolyte
- Insert new electrode
Only the mV standard 1 value is above
the mV zero value.
- Standard 1 empty
- Piston pump leaking
- New standard 1
- Replace piston & cylinder (piston
pump)
Only the mV standard 2 value is above
the mV zero value.
- Standard 2 empty
- Piston pump leaking
- New standard 2
- Replace piston & cylinder (piston
pump)
The mV zero value is in the negative
range.
- Cleaning solution empty
- Cleaning pump faulty
- Fittings leaking
- New cleaning solution
- Replace cleaning pump
- Tighten fittings
Summary of Contents for AMTAX sc
Page 2: ......
Page 4: ...4 Table of contents...
Page 12: ...12 English Operation...
Page 24: ...24 English Maintenance...
Page 32: ...32 English Troubleshooting...
Page 40: ...40 Deutsch Funktion...
Page 52: ...52 Deutsch Wartung...
Page 68: ...68 Espa ol Operaci n...
Page 80: ...80 Espa ol Mantenimiento...
Page 108: ...108 Fran ais Maintenance...
Page 124: ...124 Italiano Procedure...
Page 136: ...136 Italiano Manutenzione...
Page 152: ...152 Nederlands Bedrijf...
Page 182: ...182 Dansk Vedligeholdelse 630 35 6 265 3 570 720 35 6 265 1 1700 Table 55 Reng ringsinterval...
Page 200: ...200 Dansk Fejls gning...
Page 208: ...208 Polski U ytkowanie...
Page 218: ...218 Polski Rozwi zywanie Problem w...
Page 231: ......