4.
When the upgrade is complete, the display shows TRANSFER COMPLETE. Remove the SD
card.
5.
Restart the instrument for the upgrade to take effect.
Calibration
N O T I C E
The manufacturer recommends calibration of the analyzer after 1 day of operation to allow all system components
to stabilize.
The automatic calibration uses the installed known standards to calibrate the analyzer. Calibration
can be done manually or scheduled to occur automatically. Use the calibration menu to view
calibration data, start or cancel the auto calibration, override the auto cal settings, calibrate manually
or to reset to the default calibration.
1.
Push
cal
to display the calibration menu.
Option
Description
START MANUAL CAL
Refer to
on page 20.
SET AUTO CALIBRATION
Refer to
Schedule automatic calibrations
CALIBRATION DATA
Displays data from the last calibration and the due date and time of the next
scheduled calibration.
OVERRIDE CALIBRATION
Enter new slope and/or zero value (reagent blank). When OVERRIDE
CALIBRATION is selected, the automatic calibration is disabled for the selected
option.
OUTPUT CALIBRATION
Select a 4-20 mA output and enter the output values to be sent.
RESET DEFAULT CAL
Reset calibration data to default values and disables automatic calibration. On
completion, perform a new calibration.
Schedule automatic calibrations
1.
Push
cal
, then select SET AUTO CALIBRATION.
2.
Select SLOPE CAL or ZERO CAL.
3.
Select ACTIVATE AUTO CAL>YES.
4.
Select STD SOLUTION and enter the standard value in ppb(not applicable for ZERO CAL).
5.
Select a schedule option for calibration.
Option
Description
TIME BASE
Sets the interval between calibrations. Options: DAYS or HOURS.
WEEK DAY
Selects the day or days of the week for calibration when TIME BASE is set to DAY.
TIME
Sets the time of day for calibration when the TIME BASE is set to DAY.
SET INTERVAL
Sets the interval between automatic calibrations in hours when the TIME BASE is set to
HOURS .
Start a calibration manually
1.
Push
cal
, then select START MANUAL CAL.
2.
The type of calibration is displayed. Push
enter
to start the calibration.
20
English