5.
Select EVENT CONTROL in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
Description
SET SETPOINT
Sets the value to set the relay to on.
DEADBAND
Sets a delay so that the relay will be stable as the process value converges to the setpoint.
OnMax TIMER
Sets the maximum time the relay stays on (default = 0 min).
OffMax TIMER
Sets the maximum time the relay stays off (default = 0 min).
OnMin TIMER
Sets the time the relay stays on, independent from the measured value (default = 0 min).
OffMin TIMER
Sets the time the relay stays off, independent from the measured value (default = 0 min).
6.
Select SCHEDULER in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
Description
HOLD OUTPUTS
Holds or transfers outputs for the selected channels.
RUN DAYS
Selects the days the relay stays on. Options: Sunday, Monday, Tuesday, Wednesday,
Thursday, Friday, Saturday
START TIME
Sets the start time.
INTERVAL
Sets the time between activation cycles (default = 5 min).
DURATION
Sets the period of time the relay is set to on (default = 30 sec).
OFF DELAY
Sets the time for additional hold/output time after the relay is set to off.
7.
Select WARNING in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Option
Description
WARNING LEVEL
Sets the level for the warning activation and starts applicable individual warnings.
8.
Select PROCESS EVENT in the FUNCTION menu and then select the applicable options in the
ACTIVATION menu.
Note: More than one option can be selected.
Option
Description
MEASURING 1
Closes relay during measurement cycle.
MEASURING 2
Closes relay during measurement cycle.
MEASURING 3
Closes relay during measurement cycle.
MEASURING 4
Closes relay during measurement cycle.
MEASURING 5
Closes relay during measurement cycle.
MEASURING 6
Closes relay during measurement cycle.
ZERO CAL
Closes relay during ZERO CAL cycle.
SLOPE CAL
Closes relay during SLOPE CAL cycle.
SHUTDOWN
Closes relay when in SHUTDOWN mode.
STARTUP
Closes relay during STARTUP cycle.
GRAB SAMPLE
Closes relay during GRAB SAMPLE measurement.
MARK END OF MEASUR
Closes relay for 1 second at the end of each measurement cycle.