Option
Description
SET PARAMETER X
Selects the parameter referring to the variable X.
SET VARIABLE Y
Selects the sensor referring to the variable Y.
SET PARAMETER Y
Selects the parameter referring to the variable Y.
SET FORMULA
Selects the calculation formula to complete. Options: None, X-Y, X+Y, X/Y, [X/Y]%, [X
+Y]/2, X*Y, [X-Y]%/X
DISPLAY FORMAT
Selects the number of decimal places shown in a calculation result. Options: Auto,
XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX
SET UNITS
Enters the unit name (maximum of 5 characters).
SET PARAMETER
Enters the measurement name (maximum of 5 characters).
Configure the outputs
4–20 mA module setup
1.
Push
menu
and select SETUP SYSTEM>SETUP OUTPUTS>4–20mA SETUP.
2.
Select the OUTPUT.
3.
Select an option.
Option
Description
ACTIVATION
The menu list items change with the selected function. Refer to
on page 13 for more information.
SELECT SOURCE
Select the output. Options: None if the output is not configured, the analyzer name or
calculation if a calculation formula has been configured. Refer to
on page 12.
SET PARAMETER
Select the measurement channel from the list.
SET FUNCTION
Select a function. Further options will vary depending on which function is chosen.
LINEAR CONTROL—Signal is linearly dependent on the process value. PID CONTROL
—Signal works as a PID (Proportional, Integral, Derivative) controller. LOGARITHMIC—
Signal is represented logarithmically within the process variable range. BILINEAR—
Signal is represented as two linear segments within the process variable range.
SET TRANSFER
If TRANSFER is or will be selected as the ERROR HOLD MODE, select SET
TRANSFER and enter the transfer value. Range: 3.0 to 23.0 mA (default = 4.000). Refer
to
on page 17.
SET FILTER
Enter the filter value. This is a time-average filter value of 0 to 120 seconds (default = 0).
SCALE 0mA/4mA
Select the scale (0–20 mA or 4–20 mA).
4–20 mA activation options
1.
Push
menu
and select SETUP SYSTEM>SETUP OUTPUTS>4–20mA SETUP.
2.
Select the applicable OUTPUT.
3.
Select SET FUNCTION>LINEAR CONTROL and then select the applicable options in the
ACTIVATION menu.
Option
Description
SET LOW VALUE
Sets the low endpoint of the process variable range.
SET HIGH VALUE
Sets the high endpoint (upper value) of the process variable range.
English
13
Summary of Contents for 5500sc SiO2
Page 467: ......