![Hach 2200 PCX Basic User Manual Download Page 16](http://html1.mh-extra.com/html/hach/2200-pcx/2200-pcx_basic-user-manual_3152422016.webp)
Figure 9 Analog output wiring
1
Analog input card
5
Optional external loop power, 15-24 VDC (150 mA)
2
Jumper position (internal loop power)
6
Wires from analog output terminals (OUT0-OUT7)
3
Jumper position (external loop power)
7
Analog connections on PLC, SCADA
4
Interconnect card
Startup
Connect the power cord
Connect the power cord to an electrical outlet with protective earth ground.
Status indicator light and count display window
When the instrument is set to on, the status indicator lights on the front panel show. In operation, the
count display window shows the particle count. Refer to
Figure 10
and to
Table 1
.
16
English