7.
Push
RATIO
to turn
Ratio mode on.
8.
Put the stray light
standard in the sample
cell holder with the
triangle on the vial
aligned with the
reference mark on the
sample cell holder.
Close the cover.
9.
Read the value
when stable. Remove
the vial from the
instrument.
10.
Record the value
on the white diamond
space on the vial using
a permanent marker.
Measure the Gelex secondary turbidity standards
Measure the Gelex secondary turbidity standards each time the instrument is calibrated and record the new values on the Gelex vials with a water
soluble marker.
1.
Clean the Gelex
vials with a soft, lint-free
cloth to remove water
spots and fingerprints.
2.
Apply a small bead
of silicone oil from the
top to the bottom of the
vial.
3.
Use the oiling cloth
to apply the oil equally
to the surface of the
vial. Remove the
excess oil. Make sure
that the vial is almost
dry.
4.
Push
RANGE
to
select automatic
ranging.
"AUTO" is shown below
the Range Mode label
on the instrument.
5.
Push
SIGNAL AVG
to turn signal averaging
off.
The SIGNAL AVG light
turns off.
6.
Push
UNITS/Exit
to
select the NTU
measurement mode.
16
English
Summary of Contents for 2100AN IS
Page 1: ...DOC022 53 80206 2100AN IS 05 2014 Edition 4 User Manual...
Page 2: ......
Page 51: ......