29
Section 4
Operation
4.1 Analyse COD with LANGE Cuvette Tests
DANGER
Handling chemical samples, standards and reagents can be dangerous. Users
of this product are advised to familiarize themselves with safety procedures
and the correct use of chemicals, and to carefully read all relevant material
safety data sheets.
To analyse COD follow the working procedure enclosed in the according Cuvette Test
LCK 014/114/314/614.
4.2 Quality assurance
The measurement results must be subjected to plausibility checks (dilute and/or
spike the sample).
Regular checks with corresponding standard solutions (see
Parts and Accessories on page 31
) ensure that your measurement system is
functioning properly and is being correctly operated.
4.3 Evaluate LANGE COD Cuvette Tests
Note:
Make sure that the instrument is programmed with the corresponding calibration data,
see
3.4 Check entered data on page 28
.
1.
Preparation of a reagent blank: Fill a cuvette with the appropriate volume of DI
water instead of sample. Like for the sample cuvette accomplish exactly all
further working steps accordingly to the LCK working procedure.
Note:
Only one reagent blank per each test charge is necessary. Store it darkly.
2.
Press
POWER
to switch on the instrument. COD LCK 014/114 with POCKET
COLORIMETER II 600 nm, COD LCK 314/614 with POCKET COLORIMETER
II 450 nm.
3.
Make sure that the correct measuring range is selected!
Otherwise press
MENU
. "SEL" is displayed. A flashing arrow indicates the
current measuring range (1 or 2). Press
GREEN
key to select measuring range.
Press
MENU
to get back to the measuring mode.
4.
Insert the reagent blank into the cell holder equipped with the cuvette adapter
1 cm.
Note:
Make sure that the cuvette adapter is positioned correctly.
5.
Place the instrument cap with the HACH LANGE label over the cell holder to
cover the cuvette.
6.
Press
BLUE
key. "0" (LCK 014/114) or "0.0" (LCK 314/614) is displayed.
7.
Remove the instrument cap and insert the sample cuvette.
8.
Place the instrument cap over the cell holder to cover the cuvette.
9.
Press
GREEN
key.
Summary of Contents for POCKET COLORIMETER II Series
Page 2: ......
Page 4: ...2 ...
Page 6: ...4 ...
Page 8: ...6 ...
Page 10: ...8 ...
Page 12: ...10 ...
Page 14: ...12 ...
Page 16: ...14 ...
Page 18: ...16 ...
Page 20: ...18 ...
Page 22: ...20 ...
Page 34: ...32 ...
Page 36: ...34 ...
Page 38: ...36 ...
Page 40: ...38 ...
Page 49: ...Inbetriebnahme 47 3 BLAUE Taste drücken bis dFL angezeigt wird GRÜNE Taste drücken ...
Page 50: ...48 ...
Page 54: ...52 ...
Page 56: ...54 ...
Page 58: ...56 ...
Page 60: ...58 ...
Page 70: ...68 ...
Page 74: ...72 ...
Page 76: ...74 ...
Page 78: ...76 ...
Page 80: ...78 ...
Page 92: ...90 ...
Page 94: ...92 ...
Page 96: ...94 ...
Page 98: ...96 ...
Page 108: ...106 ...
Page 112: ...110 ...
Page 114: ...112 ...
Page 116: ...114 ...
Page 118: ...116 ...
Page 130: ...128 ...
Page 132: ...130 ...
Page 134: ...132 ...
Page 136: ...134 ...
Page 144: ...Systemstart 142 2 Tryck på BLÅ tangent för att bläddra till dFL Tryck på GRÖN tangent ...
Page 148: ...146 ...
Page 150: ...148 ...
Page 151: ......