background image

6

Tank wall mount kit, LZX414.00.71000 and LZX414.00.72000

Install the tank wall mount

1.

Find a suitable installation location and install the quick-fit anchors.

Note: 

Drill accurate holes so that the fastening lug is exactly vertical. 

a.

Drill the holes (see 

Figure 5 on page 7

) and insert the quick-fit anchors.

b.

Hit the head of the quick-fit anchor with a hammer to make the quick-fit anchor 
spread in the hole.

Note: 

Only tighten the nuts after the base and fastening lug have been assembled and aligned.

2.

Attach the base to the quick-fit anchors using four M8 nuts and washers (see 

Figure 6 

on page 7

). Tighten the nuts by hand.

3.

Attach the fastening lug to the base using four 

cap screws 

(M8 x 30 mm) and washers 

(see

Figure 7 on page 8

). Tighten the screws by hand.

4.

Align the base and fastening lug vertically and tighten all screws.

Note: 

If the wall is not exactly vertical, place shims behind the base at the top or bottom and 

then tighten.
If the fastening lug is not vertical, move the base and the fastening lug a few degrees in 
the holes.

5.

Assemble the top jointed pipe and the bottom jointed pipe as shown in 

Figure 8 on 

page 8

.

a.

Insert the hex head screws (M8 x 30 mm) through the holes, install the washers 
and tighten the M8 nuts.

Note: 

Do not overtighten the nuts. The joint should still move to prevent damage to the rotating 

scraper and the sensor.

b.

Lock the screws using M8 nuts.

6.

Loosely bolt the threaded brackets using cap screws (M5 x 20 mm) and spring 
washers to the unthreaded brackets on the jointed mount pipe (see 

Figure 9 on page 

9

). During this process pay attention to the installation dimensions (see 

Figure 3 on 

page 4

 an

Figure 4 on page 5

).

CAUTION
Risk of injury. Tighten the screws sufficiently so that the brackets cannot slide off 
the mount pipe.

7.

Based on the local conditions and the immersion depth required (see operating 
instructions for SONATAX sc), determine the installation position for the two top 
brackets and slide the mount pipe and brackets on the fastening lug.

Note: 

The flange on the bottom pipe with joint should be approximately 3–5 cm under the 

surface of the water so that the sensor always remains immersed while measurements 
are made.

Important Note: 

Align the brackets on the fastening lug and align the joint so that the joint 

axis is vertical to the movement of the rotating scraper.

8.

Check the dimensions and tighten all screws.

Summary of Contents for LZX414.00.71000

Page 1: ...DOC273 52 00144 Jul07 Tank wall mount kit LZX414 00 71000 and LZX414 00 72000 Instruction Sheet Edition 1 July 2007 wc kt ...

Page 2: ... HACH LANGE GmbH 2007 All rights reserved Printed in Germany ...

Page 3: ...s are shown in Figure 3 on page 4 and Figure 4 on page 5 Figure 1 Items supplied with tank wall mount kits LZX414 00 71000 and LZX414 00 720001 1 Mount pipe 2 m jointed Cat No BRO072 9 Hex head screw M8 x 30 mm 2x 2 Pipe 1 m jointed Cat No BRO077 only LZX414 00 71000 10 Washer for M8 2x 3 Pipe 0 35 m jointed Cat No BRO078 only LZX414 00 72000 11 Nut M8 4x 4 O ring for SONATAX sc Cat No LZY453 12 S...

Page 4: ...49 2x 30 Cap screw M8 x 30 mm 4x 19 Bracket threaded 31 Washer for M8 4x 20 Bracket unthreaded 32 Plug Cat No LZX909 21 Set of cap screws Cat No LZY445 33 Impact protector Cat No LZY364 22 Cap screw M5 x 20 mm 3x 34 Protective plate 23 Washer for M5 mm 3x 35 Spacer 2x 24 Fastening lug 36 Pipe clamp 2x 25 Set of quick fit anchors Cat No LZY428 37 Hex wrench 5 mm 26 Nut M8 4x 38 Flat head cap screw ...

Page 5: ...3 Tank wall mount kit LZX414 00 71000 and LZX414 00 72000 Figure 2 Dimensions 1 Jointed pipe 0 35 m Cat No BRO078 from LZX414 00 72000 2 Jointed pipe 1 m Cat No BRO077 from LZX414 00 71000 ...

Page 6: ...4 Tank wall mount kit LZX414 00 71000 and LZX414 00 72000 Figure 3 Installation dimensions for LZX414 00 71000 ...

Page 7: ...r manual Note When selecting the install location always make sure there is safe access to all parts of the system and that the SONATAX sc can never hit the tank wall During the selection of the install location pay attention to The load bearing capacity and strength of the tank wall lever forces due to flow and installation The maximum cable and tubing lengths Note Refer to the SONATAX sc user ma...

Page 8: ...m jointed pipe as shown in Figure 8 on page 8 a Insert the hex head screws M8 x 30 mm through the holes install the washers and tighten the M8 nuts Note Do not overtighten the nuts The joint should still move to prevent damage to the rotating scraper and the sensor b Lock the screws using M8 nuts 6 Loosely bolt the threaded brackets using cap screws M5 x 20 mm and spring washers to the unthreaded ...

Page 9: ...414 00 72000 9 To attach the sensor remove the jointed mount pipe from the fastening lug see section Attach the SONATAX sc on page 9 Figure 5 Install the quick fit anchors Figure 6 Attach the base 1 Nut M8 3 Base 2 Washer for M8 4 Quick fit anchor ...

Page 10: ...re 7 Attach the fastening lug 1 Cap screw M8 x 30 mm 4 Quick fit anchor 2 Washer 5 Base 3 Fastening lug Figure 8 Assemble the joint 1 Parts of joint jointed mount pipe 4 Nut M8 2 Parts of joint jointed pipe 5 Washer for M8 3 Lock nut M8 6 Hex head screw M8 x 30 mm ...

Page 11: ...e and use them to seal the short bottom jointed pipe at the joint see Figure 11 on page 10 3 Slide a plug and sealing plug over the cable to seal the long top jointed mount pipe at the joint 4 Slide the sensor cable through the long top jointed mount pipe 5 Using the plug and sealing plug seal the long top jointed mount pipe at the joint 6 Insert a plug and a sealing plug in the top of the mount p...

Page 12: ... 00 71000 and LZX414 00 72000 Figure 10 Attach the SONATAX sc 1 Sensor cable 4 Washer 2 Bottom jointed pipe 5 O ring 3 Cap screw M5 x 20 mm 6 SONATAX sc Figure 11 Insert the plugs 1 Sealing plug 3 Sensor cable 2 Plug 4 Jointed pipe ...

Page 13: ...nd washers 2 Move the protective plate so that the sensor is suspended approximately 5 cm above the bottom edge of the protective plate Important Note Align the protective plate so that it is vertical to the movement of the rotating scraper and parallel to the axis of the joint Make sure the spacers do not block the probe LED and the probe is aligned with 2 feet towards the plate 3 Tighten the nut...

Page 14: ...ll the parts taking into account the installation dimensions and the local conditions Figure 13 on page 12 Check the position of the sensor in the tank and tighten all screws 2 Route the sensor cable to the controller and attach the quick connect fitting 3 Continue as described in the operating instructions for the SONATAX sc and the controller Figure 13 Tank wall mount assembled without sensor ...

Page 15: ...e 2 m jointed 1 BRO072 Pipe 1 m jointed for LZX414 00 71000 1 BRO077 Pipe 0 35 m jointed for LZX414 00 72000 1 BRO078 Plug 1 LZX909 Sealing plug 1 LZX908 O ring 1 LZY453 Set of accessories joint without stop 1 LZX764 Set of cap screws M5 x 12 mm 1 LZY362 Set of cap screws M5 x 20 mm 1 LZY445 Set of cap screws M8 x 30 mm 1 LZY429 Tank wall mount SONATAX sc 1 LZY370 Set of quick fit anchors 1 LZY428...

Page 16: ...lange nl www hach lange nl HACH LANGE APS Åkandevej 21 DK 2700 Brønshøj Tel 45 36 77 29 11 Fax 45 36 77 49 11 info hach lange dk www hach lange dk HACH LANGE AB Vinthundsvägen 159A SE 128 62 Sköndal Tel 46 0 8 7 98 05 00 Fax 46 0 8 7 98 05 30 info hach lange se www hach lange se HACH LANGE S R L Via Riccione 14 I 20156 Milano Tel 39 02 39 23 14 1 Fax 39 02 39 23 14 39 info hach lange it www hach l...