Error codes
shows the codes that can occur for various reasons. Error codes show instrument
malfunction or user error.
Table 2 Error code descriptions
Error
Code
Description
Solution
E02
The ISFET pH probe is damaged.
Replace the probe. Contact technical support for probe
replacement information.
E03
Clean the probe.
Clean the probe. If the error continues, replace the probe.
E04
Glass pH probe slope error. The
slope is smaller than 85% or larger
than 102% of 59.16 mV per pH unit.
Clean the probe. If the error continues, replace the probe.
E06
ISFET pH probe slope error.
Clean the probe. Soak in 40 ºC (113 ºF) pH 7.00 buffer for
2 minutes. Calibrate again. If the error continues, replace the
probe.
E07
Clean the probe.
Clean the probe. Make sure that no air bubbles on the sensor
surface or foreign objects or materials on the sensor.
E08
Too long to calibrate. Signal is not
stable during calibration.
Clean the probe. Disconnect stirrers and other AC power
sources. Make sure that the calibration solution temperature
is constant. Replace the probe if the error continues.
E13
The ISFET pH probe temperature
sensor is damaged.
Replace the probe. Contact technical support for probe
replacement information.
E14
The battery is extremely low.
Replace the batteries immediately. Damage to the accuracy
and function of the meter could have occurred.
E15
Replace the battery immediately.
The accuracy and function of the
meter could be compromised.
Contact technical support for service information.
E20
The conductivity temperature
sensor is damaged.
Replace the probe. Contact technical support for probe
replacement information.
E25
The slope is smaller than 60% or
larger than 140% of the nominal.
Replace DO probe membrane and fill solution. Replace the
probe if the error continues. Nominal = 0.1822 ppm/mV.
E26
The mV reading is more than
±10 mV from nominal.
Replace DO probe membrane and fill solution. Replace the
probe if the error continues. Nominal = 0.0 mV at 0%
saturation; 45mV at 100%.
E27
DO probe temperature error
Attach the temperature sensor to the 3.5 mm phono jack. DO
readings are highly temperature dependent and a
temperature sensor must be attached.
E28
Barometric pressure sensor error
Contact technical support for repair information.
E30
ISE electrode calibration error
The slopes are not the same sign or are not within 25% of
each other. Calibrate again in the correct solutions.
E40
Unrecognized host command
Command from a host PC is not found. Use only valid
commands.
E42
Invalid input
The value entered during the setup is invalid. Enter a
different value.
E44
No probe is installed
No probe is installed for the applicable parameter. Set the
meter power to off. Install the correct probe. Set the meter
power to on.