9.3 Cleaning and disinfection
The Haag-Streit slit lamps and their accessories can, if required, be carefully wiped
down with ready-for-use disposable 70% ethanol disinfectant wipes. Surface-friendly
disinfectants (containing aldehyde or aldehyde-free) are also permitted, such as
Kohrsolin FF.
WARNING!
• The preparation instructions provided do not apply to tonometer
measuring prisms.
• Tonometer measuring prisms must be prepared in accordance with
a different manual.
• Too strong or aggressive disinfectants or cleaning liquids e.g.
hydrogen peroxide will damage the finish and coating of the device.
• Do not use sprays.
• Observe the manufacturer's safety instructions.
• Do not use any cloths that drip.
• Wring out any soaked cloths before use when necessary.
• Ensure that no liquid penetrates the device.
• Comply with the stipulated exposure time.
• Clean optical surfaces after disinfection with a very soft cloth.
NOTE!
IP code: IPX0 (device is not protected against liquids)
10 Appendix
10.1 Accessories / functionals parts / detachable parts /
consumables