10.6.3 Electromagnetic immunity environment tested (part 1)
This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product should assure that it is used in such an envir-
onment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment Ð guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV, 100kHz for power supply lines *
± 1 kV, 100 kHz for input/output lines *
± 2 kV, 100kHz for power supply lines *
± 0.5, ± 1 kV, 100 kHz for input/output
lines *
Mains power quality should be that of a typical com-
mercial or hospital environment.
* Not applicable for DC and I/O if cable < 3 m.
Surge
IEC 61000-4-5
± 0.5, ± 1 kV line(s) to line(s) *
± 0.5, ± 1, ± 2 kV line(s) to earth *
± 1 kV line(s) to line(s) *
± 0.5, ± 1, ± 2 kV line(s) to earth *
Mains power quality should be that of a typical com-
mercial or hospital environment.
* Not applicable for DC and I/O if cable < 3 m.
Voltage dips, short interruptions
and voltage variations on power
supply input lines
IEC 61000-4-11
0% U
T
: 0.5 cycle at 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0% U
T
: 1 cycle at 0°
0% U
T
: 250/300 cycles at 0°
70% U
T
: 25/30 cycles at 0°
0% U
T
: 0.5 cycle at 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0% U
T
: 1 cycle at 0°
0% U
T
: 250/300 cycles at 0°
70% U
T
: 25/30 cycles at 0°
Mains power quality should be that of a typical com-
mercial or hospital environment.
If the user of this product requires continued opera-
tion during power mains interruptions, it is recom-
mended that this product be powered from an unin-
terruptible power supply or battery.
U
T
is the a.c. mains voltage (100 – 240 V) prior to
application of the test level.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50/60 Hz
30 A/m
50/60 Hz
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical com-
mercial or hospital environment.