9 Maintenance
WARNING!
• Do not modify this device without authorization of the manufacturer.
Installation and repairs may only be performed by trained
specialists.
• Contact your Haag-Streit representative for installation, repairs and
modification work on the system. The contact details are available at
www.haag-streit.com.
• Only original Haag-Streit spare parts may be used.
9.1 Device inspection
To check the correct function, proceed as follows:
• Insert the test rod into the radial movement bearing, whilst at the same time
aligning the surface to the microscope at a right angle.
• Set the slit length to 8 or 14 mm.
• Set the illumination intensity to 50%.
• Set the magnification in the microscope to max.
• Set the eyepieces in such a way that the test rod is in sharp focus. In doing so,
turn the ocular from the (+) to the (-) side.
• Switch on the camera.
• For all magnifications, the structure of the test rod in the ocular and in the
camera image must be shown in sharp focus.
• Turn on the illumination of the 0.2 mm diaphragm.
• Turn on the crosshairs in EyeSuite; the cross must be inside the illuminated dot.
9.2 Servicing
To guarantee a long service life, the device must be cleaned weekly as described
and protected with the dust cover when not in use. We recommend having the
device inspected once a year by an authorized service technician.
9.3 Cleaning and disinfection
The Haag-Streit slit lamps and their accessories can, if required, be carefully wiped
down with ready-for-use disposable 70% ethanol disinfectant wipes. Surface-friendly
disinfectants (containing aldehyde or aldehyde-free) are also permitted, such as
Kohrsolin FF.
WARNING!
• The preparation instructions provided do not apply to tonometer
measuring prisms.
• Tonometer measuring prisms must be prepared in accordance with
a different manual
• Too strong or aggressive disinfectants or cleaning liquids e.g.
hydrogen peroxide will damage the finish and coating of the device.
• Do not use sprays
• Observe the manufacturer's safety instructions
• Do not use any cloths that drip
• Wring out any soaked cloths before use when necessary
• Ensure that no liquid penetrates the device
• Comply with the stipulated exposure time
• Clean optical surfaces after disinfection with a very soft cloth
NOTE!
IP code: IPX0 (device is not protected against liquids)
10 Appendix
10.1 Accessories / functionals parts / detachable parts /
consumables
Components
REF
Set white balance grey card
1021485
Computer cable USB 3.0
1022373