14
DEUTSCH
ENGLISH
FRANÇAIS
ITALIANO
ESPAÑOL
NEDERLANDS
PORTUGUÊS
SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. no. 1500.7220551‑04070 – 2021 – 08
6.12 Error messages (illumination head)
1)
2)
3)
Background
illumination
Polarity reversal
Slit lamp
illumination
ERROR
Error messages
Measures
E1
Incorrect supply polarisation
Contact your Haag-Streit representative.
x
Red
x
E2
Illumination control not
recognized
Connect illumination control or replace, if necessary.
Red
x
Red
E3
Temperature is too high
The light sources’ power will be reduced. Normal operation is
ensured once the permissible temperature has been reached.
Red, flashing
x
Red, flashing
E4
No communication between
power supply and illumination
Contact your Haag-Streit representative.
Red, flashing 2 ×
x
Red, flashing 2 ×
E6
General error
Send PS-LED to the appropriate service branch.
Red, flashing 4 ×
x
Red, flashing 4 ×
6.13 Error messages release module RM02
ERROR
Error messages
Measures
Operational control LED (75)
E14
No communication with LED illumination LI01 plus / LI02 plus
Contact your Haag-Streit representative.
Red, flashing 2 ×
E16
General error
Send device to the appropriate service branch.
Red, flashing 4 ×
6.14 Error messages camera module CM03
ERROR
Error messages
Measures
Operational control LED (26)
E18
No communication with LED illumination LI01 plus / LI02 plus
Contact your Haag-Streit representative.
Red