NOTE!

This equipment must only be operated by qualified personnel. The 

owner is responsible for their training. This device may only be used 

in accordance with the instructions in "Purpose of use".

1.6  Light toxicity

WARNING!

As extended, intensive illumination can damage the retina, the use of the de- 

vice in the examination of the eye should not be prolonged unnecessarily. The 

illumination of this slit lamp emits a radiation in the range between 400 and 

750nm. The retinal dose for a photochemical risk is composed of the product 

of the radiance and the exposure time. If the radiance is halved, the time until 

the exposure time limit value is reached will double accordingly. To date, no 

acute, optical radiation hazard has been detected in slit lamps. Nevertheless, 

we recommend keeping the intensity of the light reaching the patient's retina 

to the minimum possible for the respective diagnosis. Children, people with 

aphakia and people suffering from eye conditions are most at risk. An increas- 

ed risk may also occur if the retina is exposed to the same or a similar device 

with a visible light source within 24 hours. This applies, in particular, if the 

retina has been photographed with a flashbulb in advance. The light from this 

instrument may be dangerous. The risk of eye damage increases with the 

exposure time. An exposure time with this instrument at maximum intensity of 

longer than 143 seconds exceeds the guideline value for a risk.

1.7 Disinfection

NOTE!

The device does not need to be disinfected. For more information on

cleaning, please refer to the 'Maintenance' section.

1.8  Warranty and product liability

Haag-Streit products must be used only for the purposes and in the manner des-

cribed in the documents distributed with the product. 

The product must be treated as described in the ‘Safety’ chapter. Improper hand-

ling can damage the product. This would void all guarantee claims. 

Continued use of a product damaged by incorrect handling may lead to personal 

injury. In such a case, the manufacturer will not accept any liability.

Haag-Streit does not grant any warranties, either expressed or implied, including 

implied warranties of merchantability or fitness for a particular use.

Haag-Streit expressly disclaims liability for incidental or consequential damage 

resulting from the use of the product.

This product is covered by a limited warranty granted by your seller.

This product is covered by a limited warranty, which may be reviewed at 

www.haag-streit-usa.com.




For USA only:

Haag-Streit products must be used only for the purposes and in the manner des-

cribed in the documents distributed with the product. 

The product must be treated as described in the ‘Safety’ chapter. Improper hand-

ling can damage the product. This would void all guarantee claims. 

Continued use of a product damaged by incorrect handling may lead to personal 

injury. In such a case, the manufacturer will not accept any liability.

Haag-Streit does not grant any warranties, either expressed or implied, including 

implied warranties of merchantability or fitness for a particular use.

Haag-Streit expressly disclaims liability for incidental or consequential damage 

resulting from the use of the product.

This product is covered by a limited warranty granted by your seller.

This product is covered by a limited warranty, which may be reviewed at 

www.haag-streit-usa.com.




For USA only:

1.9 Symbols

Read the instructions for use

attentively

Notes on disposal, see the

'Disposal' chapter

Year of production

Manufacturer

Serial number

HS reference number

Test symbol of CSA with ap-

proval for USA

European certificate of 

conformity

Testsymbol of TÜV Rheinland 

with approval for INMETRO 

Brasil 

2. Introduction

The slit lamp comprises an illumination and a binocular microscope. The entire de-

vice can be moved in front of the eyes using the instrument base. The illumination 

offers a range of setting possibilities for making the practically invisible areas of the 

eye visible. There is also a range of accessories available for the slit lamp to allow 

special diagnosis possibilities in addition to the general examinations.

01_IFU_BM900-7200492-04150_eng.indd   5

21.02.2017   09:54:55

Summary of Contents for BM 900

Page 1: ...00 7200492 04150 15 Edition 2017 01 DEUTSCH ENGLISH FRAN AIS ITALIANO ESPA OL NEDERLANDS PORTUGU S SVENSKA DOC no 1500 1500 1400209 04000 INSTRUCTIONS FOR USE Slit lamp BM 900 15 Edition 2017 01 01_IF...

Page 2: ...of diseases or trauma which affect the structural properties of the eye Contraindication There is no absolute contraindication for tests with this device Appropriate profes sional judgement and cautio...

Page 3: ...Preparing the patient 7 5 3 Operating the instrument 7 5 4 Setting the filters diaphragms 8 5 5 Fixation star 9 5 6 Microscope and eyepiece 9 6 Decommissioning 9 7 Technical data 9 7 1 Slit illuminat...

Page 4: ...npack the appliance check whether the packaging shows traces of incorrect handling or damage If this is the case notify the transport company that has delivered the goods to you Unpack the equipment t...

Page 5: ...nued use of a product damaged by incorrect handling may lead to personal injury In such a case the manufacturer will not accept any liability Haag Streit does not grant any warranties either expressed...

Page 6: ...Breath shield option 14 Protective cover 15 Illumination unit microscope an gle scale 16 Illumination arm locking screw 17 Microscope arm locking screw 18 Slit width setting screw 19 Slit length diap...

Page 7: ...s and press the rocker switch The green light on the rocker switch illuminates when the device is switched on Turn the rotating knob on the illumination control to a position between 1 and 10 5 Operat...

Page 8: ...d 10 a short mirror 12 is used the illumination turned 90 using the locking screw 20 and tilted in 5 steps us ing the latch 27 and the illumination and microscope turned to the central po sition 0 The...

Page 9: ...Dioptre scale for 16x eyepieces for setting the refraction of the examin er 8 D 46 Dioptre scale for 10x eyepieces for setting the refraction of the examin er 8 D 47 Index white point 48 10x or 25x e...

Page 10: ...from the to the side The structure of the test rod must be in sharp focus in all magnifications Close slit edges to approx 0 5 mm The borders must be in sharp focus here Completely open slit edges and...

Page 11: ...15004 2 standards into account The EN 60601 1 standard must be observed when using different medical and or non medical electrical devices in combination Compliance of the BM 900 slit lamp with the Di...

Page 12: ...sistance to interfe rence Third party devices may only be connected in compliance with the EN 60601 1 standard F 2 Emitted interference standard table 1 The information is based on the requirements of...

Page 13: ...or hospital environment Surge EN 61000 4 5 1 kV for symmetrical voltages 2 kV for asymmetrical voltages 1 kV for symmetrical voltages 2 kV for asymmetrical voltages Mains power quality should be that...

Page 14: ...ded separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interf...

Page 15: ...minal output of the transmitter W Safe distance according to transmission frequency m 150 kHz 80 MHz D 0 35 80 MHz 800 MHz D 0 7 800 MHz 2 7 GHz D 1 4 0 01 0 1 1 10 100 For transmitters with a nominal...

Page 16: ...Should you have any further questions please contact your HAAG STREIT dealer at http www haag streit com contact contact your distributor html HAAG STREIT AG Gartenstadtstrasse 10 3098 Koeniz Switzerl...